149
_____________________________________________________________________
8.8 Internal Trigger
______________________________________________________________
Flashing cursor
OFF
Trigger filter is disabled.
ON
In Recorder function mode
Trigger filter is enabled. Filter width: 10 ms
0.1 to 10
In Memory recorder function
Trigger filter is enabled. Filter width (DIV)
0 V
-600 mV
B
0 V
A
Point A
trigger level: 200 mV,
trigger direction (slope): rising
( )
Point B
trigger level: -600 mV,
trigger direction (slope): falling
( )
200 mV
5. Move the flashing cursor to the position shown in the
figure below, and set the trigger filter width using the
buttons. (Filter settings are not available for the RMS
Recorder.)
To cause triggering at point A or point B with the sine
wave shown below, make the following settings.