47
_____________________________________________________________________
4.4 Using the X-Y Waveform Plots (MEM)
______________________________________________________________
Flashing cursor
4
5
4.4 Using the X-Y Waveform Plots (MEM)
X-Y combined plotting can be performed when the Display
Layout is set to X-Y display on the Status Screen. For the
8808-01, 8808-51 when the X axis is assigned to one
analog channel, up to three X-Y combined plots are
possible.
Voltage axis magnification/compression is active also when
using X-Y combined plotting.
Using the A/B cursors, it is possible to specify the data
between the cursors for partial plotting.
Procedure
Status/ Channel Screen
1. Press the
SET
button to display the Status Screen.
2. Move the flashing cursor to
Display Format
and select format
(
X-Y Dot
or
X-Y Line
) using the
buttons.
3. Press the
SET
button with the Status Screen displayed to
display the Channel Screen.
4. Select the display color using the
buttons.
5. Move the flashing cursor to the channel to use as the X
axis, and select it using the
buttons.
Channels other than the one selected for the X axis are the
Y-axis channels.