7.3 Setting the Channel Screen
______________________________________________________________
Flashing cursor
Waveform display
area (upper and
lower limit)
NOTE
The full scale of the product's analog input unit is 10 DIV times
the voltage-axis range.
On the other hand, the actual range of measurement for the
voltage range is about
1.2 times the full scale of the voltage
axis. For example, on the 10 mV range, 10 mV x 10 DIV =
100 mV full scale, and 1.2 times that value is the approximate
measurable area:
120 mV.
Measurements that fall outside the measurable range are
indicated as approximately
1.2 times full scale.
7.3.3 Setting the Voltage Axis Range
The voltage axis range for each channel is set. The set value
denotes the voltage value for 1 division along the
measurement range (vertically).
Procedure
Channel Screen
1. Press the
SET
button to display the Channel Screen.
2. Move the flashing cursor to the desired channel position.
3. Use the
buttons to make a setting.