Fan Filter Replacement
Exicor GEN5 User Manual
69
Figure 6.2 DC Level Indicator
Compare these values to the values before shipment of the system (provided on the
ATP document). If either of these values falls below 1.5V, the laser and/or detector
filters need to be changed. To determine if the decreased DC level is caused by the
laser or detector filters, measure the laser power at the stage and compare it to the
power stated on the ATP document. The laser should be replaced when the power
falls below 1.5 mW. If the laser power is still above 1.5 mW then the detector filters
are the probable cause of low DC levels.
Fuses
The Exicor GEN5 contains no user replaceable fuses. In the event of a system
power problem, contact Hinds Instruments
.
Fan Filter Replacement
The Exicor GEN5 has one cooling fan to control and stabilize the temperature of the
system electronics. The inlet has a filter to keep dust and other particles out of the
system. The filter is located at the front of the control system on the left side of the
Master Electrical Enclosure (see the following figure).
Summary of Contents for Exicor Gen5
Page 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B ...
Page 2: ......
Page 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B Software 025 0000 066 Version 1 0 ...
Page 10: ...Table of Tables viii Exicor GEN5 User Manual ...
Page 22: ...Operators Safety Summary 12 Exicor GEN5 User Manual ...
Page 72: ...Chapter 4 Software Interface 62 Exicor GEN5 User Manual ...
Page 90: ...Chapter 7 Troubleshooting Help 80 Exicor GEN5 User Manual ...
Page 94: ...Appendix A Technical Specifications 84 Exicor GEN5 User Manual ...
Page 97: ......