Overview of the Graphical User Interface
Exicor GEN5 User Manual
35
Birefringence Plot Area
The birefringence plot area shows the currently active measurement data or
currently open data file.
Using a mouse, you can select a grid point to display the measured data in the
Interactive Scan panel. If you select more than one grid point (by dragging the
mouse), you can either outline an area for a new scan or you can zoom in on the
area with the
button. The coordinates of the selected area are displayed in the
Automatic Scan panel.
The
button allows you to zoom in on a grid point or points. The
button
allows you to zoom in on the data from a completed scan. Click the
button to
show the entire scannable area. Alternatively the mouse wheel can be used for
zooming in and out.
By double-clicking on a grid point, you command the system to move to that location
and take a measurement (the same as clicking once and then clicking the Move &
Sample button). A cross-hair which represents the current stage position is
displayed.
The default grid spacing is set at 50mm. If necessary, you can change the grid
spacing and grid units. See “Changing the Grid Spacing” and “Changing the Stage
Distance Units” on page 39 for more information.
Other Displays and Controls
Figure 4.10 shows the real-time signal DC portion of the detector signal. The DC
level is important because it represents the light intensity that is reaching the
detector. A low DC level means the light is being blocked in some way. More light
reaching the detector means more signal strength and less noise in the
measurement. See the Troubleshooting chapter for more information.
Figure 4.10 DC Level Indicator
Figure 4.11 shows the
Offsets
,
Notes
, and
Home & Reset
buttons in the lower-left
area of the Exicor window..
Summary of Contents for Exicor Gen5
Page 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B ...
Page 2: ......
Page 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B Software 025 0000 066 Version 1 0 ...
Page 10: ...Table of Tables viii Exicor GEN5 User Manual ...
Page 22: ...Operators Safety Summary 12 Exicor GEN5 User Manual ...
Page 72: ...Chapter 4 Software Interface 62 Exicor GEN5 User Manual ...
Page 90: ...Chapter 7 Troubleshooting Help 80 Exicor GEN5 User Manual ...
Page 94: ...Appendix A Technical Specifications 84 Exicor GEN5 User Manual ...
Page 97: ......