Emergency OFF Switch
Exicor GEN5 User Manual
9
Emergency OFF Switch
The Exicor GEN5 has 7 Emergency OFF (EMO) switches. Each switch consists of a
large red button on a yellow background. These switches disconnect power to all
system components except the computer and monitor. Figure S.6 shows the
location of the Emergency OFF switches.
Figure S.6 EMO Switch Locations
Activating the EMO System
In the event that an unsafe event or condition is observed with the system, an EMO
switch should be activated by pushing or slapping the red button. Pressing any
EMO button will remove power to all system components of the Sample Scanning
Module. Activating the EMO system will not harm the Exicor GEN5.
Activating the EMO system
will not
remove power to the Exicor GEN5 PC, monitor,
or third party equipment located at the user station. The power cord acts as the
emergency disconnect device for the computer and monitor.