Opening Covers
Exicor 150AT User Manual
69
Opening Covers
Check periodically and whenever the system is moved or maintained, that the
covers on the scanning module are secured and tight fitting. Be certain that no
unwanted laser light escapes from the enclosure.
Check periodically that interconnecting cables are free of sharp bends and cuts or
abrasions.
For safety, the instrument housings should be opened only by trained and
authorized maintenance personnel.
Persons changing the cables or opening the covers on the system must:
•
Remove power to the system.
•
Understand the purpose, connections and voltages at each cable connection.
•
Understand the hazards associated with the laser light used.
•
Not allow eye exposure directly from laser light or any reflection of laser light
from a mirror like surface.
•
Understand the operation of the laser high-voltage connection and the
associated shock hazard.
Cleaning the Optics
Under normal operation, the optics do not require cleaning. Never clean or touch
the optics when the system power is ON.
If a substance such as Index Matching fluid (coupling oil) drops into the detector
aperture, the oil could coat the second PEM and cause a loss of signal at the
detectors. If this occurs, clean the PEM with cotton swabs moistened with solvent.
Wipe the surface gently while rotating the swab slowly. This action ensures that the
contaminants are lifted away from the surface of the optic. See figure 5.2.
Direction of
wiping motion
Figure 5.4 Direction of cotton swab wiping motion
Choose a solvent that will not soften or remove the locking indicator used to secure
components. Isopropyl alcohol and mineral solvent are examples of suitable
materials. Acetone, MEK or Toluol are unsuitable.
Summary of Contents for EXICOR 150AT
Page 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F ...
Page 2: ......
Page 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F Software Version 6 5 ...
Page 7: ...Table Of Contents Exicor 150AT User Manual v ...
Page 16: ...Operator s Safety Summary Exicor 150AT User Manual 6 Figure S 1 Laser Beam Path ...
Page 17: ...Laser Safety Exicor 150AT User Manual 7 Figure S 2 Compliance Labeling ...
Page 21: ...Exicor 150AT User Manual 11 ...
Page 26: ...Chapter 1 Overview of the Exicor Birefringence Measurement System 16 Exicor 150AT User Manual ...
Page 29: ...Loading a Sample Exicor 150AT User Manual 19 ...
Page 74: ...Chapter 4 Operating the System 64 Exicor 150AT User Manual ...
Page 81: ...Cleaning the Cabinet Exicor 150AT User Manual 71 ...
Page 90: ...Appendix F Exicor 150AT Limited Warranty 80 Exicor 150AT User Manual ...