Figure 5.3 X-Rail & Y-Rail Lubrication
Ball Screws
Lubricate the two ball screws once per year or every 3000 hours of use. Lubrication
schedule may vary depending on frequency of use.
Wipe screw down with a clean cloth. Move the carriage to one end of the rail. Apply
grease utilizing the grease fittings located on the ends of the carriage until grease
can be seen seeping out of both sides of the carriage near the ball screw. Move the
carriage from one end of the rail to the other. Repeat this process as necessary until
the entire length of the groove in the ball screw is lightly covered with grease.
•