Exicor 150AT User Manual
18
Emergency Switch
The red mushroom switch on the side of the scanning module serves as the
EMERGENCY STOP or FF SWITCH (EMO). Pushing or striking this switch IN shuts
off all power to the Exicor 150AT. This action will turn OFF the laser.
To restore power, pull the red button OUT.
Note:
The computer and associated monitor are powered
independently of the System Scanning Module. You must
switch the computer ON when starting the system. Striking the
EMO will not turn the PC or monitor off.
You may choose to leave the computer running even when the
Exicor 150AT System Scanning Module Switch is turned off.
Fuse Types
The Exicor 150AT instrument has a total of three fuses. All three fuses are located
at the back panel of the System Scanning Module. Two of these fusses are located
at the main power input and the third is located at the laser fuse holder.
Main System Power Fuse (Qty 2)
Voltage
Fuse Rating
Hinds Part Number
ALL
5x20mm, 3.15A, 250V, SB 345-8000-034
Table 2.1 Main System Power Fuse
Laser Fuse
Voltage
Fuse Rating
Hinds Part Number
All
3.AG, T 250V/3.0A SB
345-8000-026
Table 2.2 Laser Fuse
Proper System Airflow
Airflow in the System Scanning Module enters through a filter element on the case
back and exits through a fan on the case top. The airflow helps maintain a lower
ambient temperature inside the case extending the life of the interior components.
Summary of Contents for EXICOR 150AT
Page 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F ...
Page 2: ......
Page 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 001 UM Rev F Software Version 6 5 ...
Page 7: ...Table Of Contents Exicor 150AT User Manual v ...
Page 16: ...Operator s Safety Summary Exicor 150AT User Manual 6 Figure S 1 Laser Beam Path ...
Page 17: ...Laser Safety Exicor 150AT User Manual 7 Figure S 2 Compliance Labeling ...
Page 21: ...Exicor 150AT User Manual 11 ...
Page 26: ...Chapter 1 Overview of the Exicor Birefringence Measurement System 16 Exicor 150AT User Manual ...
Page 29: ...Loading a Sample Exicor 150AT User Manual 19 ...
Page 74: ...Chapter 4 Operating the System 64 Exicor 150AT User Manual ...
Page 81: ...Cleaning the Cabinet Exicor 150AT User Manual 71 ...
Page 90: ...Appendix F Exicor 150AT Limited Warranty 80 Exicor 150AT User Manual ...