EN
24/104
1
Use according to specification
The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG.
The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also
includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this
purpose. A different use or application over and above this is deemed not in accordance with the specifications. The
company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
2
Residual risks
The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3
Technical specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model
ROTANTA 460
Type
5600
5600-01
Mains voltage (
r
10%)
208-240 V 1
a
110-127 V 1~
Mains frequency
50 - 60 Hz
50 - 60 Hz
Connected load
1500 VA
1300 VA
Current consumption
6.5 A
11 A
Cooling medium
----
Max. capacity
4 x 750 ml
Allowed density
1.2 kg/dm
3
Speed (RPM)
15000
Force (RCF)
24400
Kinetic energy
41000 Nm
Obligatory inspection (BGR 261)
yes
Ambient conditions (EN 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 40°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
,,
Pollution degree
2
Device protection class
,
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference (suppression
of radio interference)
EN 55011,
Group 1, Class B
EN 61000-3-3
FCC Class A
Interference immunity
EN 61000-6-1
----
Noise level (dependent on rotor)
d
67 dB(A)
Dimensions
Width
554 mm
Depth
707 mm
Height
453 mm
Weight
106 kg
115 kg