Fig. 162: Passive dehumidification
ventilation opening
1 Ventilation opening for passive
dehumidification
Fig. 163: Control contact
energy, low temperatures might no longer be able to be reached. The
lowest temperature that can still be reached can be found in the heat
compensation diagram
Chapter 6.14 ‘Heat compensation’
Configure the external device such that it starts
when the supply voltage is switched on. It is impor-
tant that the external device starts when the supply
voltage is switched on.
6.13.6.2
Passive dehumidification option
Devices with the passive dehumidification option are equipped with a
ventilation opening, which can be opened and closed via the control
contact.
NOTICE
Dehumidification is only possible if the moisture con-
tent of the air outside of the incubator is lower than
that inside the incubator.
The ambient temperature should be at least 5 °C
lower than the temperature to which the incubator is
to be heated.
If the control contact is activated here, the function is available in
manual and program mode. This means:
■
In manual mode, an additional symbol appears on the main
screen, via which the control contact can be switched on or off at
any time
Further information on page 54.
■
In program mode, there is an additional function available,
‘Con-
trol contact On/Off’, for creating programs in the editing view of
the program segments
■
The control contact can be used in the start settings for an
existing program
‘Control contact (option)’ on page 88.
If the control contact is activated in manual or program mode, the
ventilation opening is opened. If the control contact is deactivated in
manual or program mode, the ventilation opening is closed. An effec-
tive method of using the passive dehumidification option is to open
the ventilation opening during the heating-up phase. It is recom-
mended to plan at least 5 h Es for a heating-up phase from 4 °C to
37 °C.
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...