Operation
Fig. 50: Screw for separate blocking of
control panel
1 Screw
1
Fig. 51: Control panel blocking
1 Bore with thread,
[Control panel
blocking]
6.4
Fixing the mechanical display protection of the control panel
To prevent the incubator from being operated by unauthorized per-
sonnel, the mechanical display protection can be additionally
blocked.
1.
In the delivery condition, there is a screw on
the rear side of the incubator, which can be
used to block the mechanical display protec-
tion.
Remove this screw.
2.
Push the mechanical display protection to the right.
3.
Screw the screw into the
[Control panel blocking] bore.
The mechanical display protection is blocked. The door can
still be opened and closed.
6.5
Standard check before every use
Every time before using the incubator, make sure that
■
The mains voltage agrees with the specification on the name
plate.
■
The incubator is standing flat on its feet.
■
There is at least one slide-in module in the incubator.
■
The door closes tightly.
■
No alarm or error messages are pending.
■
The ventilation slots of the incubator are free.
■
The incubator has no obvious damage.
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...