en
12.2020 / Rev. 02
AB66005
132 / 174
Troubleshooting
1
[Logbook] button
2
[Close] button
8.2.2
Event overview
The events which have occurred while the program/manual mode
were running are shown collectively in a blue
‘Status information’ bar.
If you tap on the
[Logbook] button, an overview of the events, such
as the door opening/closing, program pause, program continuation,
are displayed with the start time and duration.
If this field is closed, there is the option of opening
the event overview via ‘Device settings
Logbook’
Chapter 6.13.16 ‘Logbook’ on page 114
If you tap on the
[Close] button, the event overview disappears auto-
matically. The door openings and events are divided into program
mode and manual mode.
After a preset time, the event overview disappears
automatically.
8.2.3
Tolerance band alarm
If the actual temperature exceeds or falls short of the temperature
setpoint by the set values, an alarm sounds.
The values can be set under
‘Settings
Tolerance band.’ , see
Chapter 6.13.4 ‘Tolerance band’ on page 97.
8.2.4
Temperature protection, class 3.1 / 3.2
An alarm is triggered when the temperature of the class 3.1. temper-
ature guard is exceeded or the temperature of the class 3.2. temper-
ature guard is fallen short of. The values are adjustable; see
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...