Handle only at Static Safe
Work Stations
The assemblies handled in this procedure are VERY SENSITIVE TO
DAMAGE BY STATIC ELECTRICITY.
PLEASE PAY ATTENTION TO THESE INSTRUCTIONS! Expensive assem-
blies are destroyed EVERY DAY because anti-static precautions are not
taken seriously.
WEAR AN ANTI-STATIC WRIST STRAP THAT IS CONNECTED TO EARTH
GROUND. If you ground the strap to the instrument chassis, make sure
the instrument is plugged in, otherwise you are not connected to earth
ground. An ungrounded instrument can not provide sufficient protection,
and can even build up its own static charge.
Table 5-4. Required Tools and Supplies
Tools Used On HP Part Number
Anti-static wrist strap all components 9300-1257 (Medium)
9300-0970 (Large)
Anti-static mat
9300-0797
Small pozidriv (1 pt.)
8710-0899
Medium pozidriv (2 pt.)
8710-0900
torque wrench, 72 lb-in coupler MTB 100 72 LB-IN
6” torque wrench, 10 lb-in
Clip lead capacitors purchase locally
6” nut driver connector repair
torque wrench, 25 lb-in connector repair 8710-1581
Microwave Connector Care Manual all connectors 08510-90064
3.5 mm connector
3.5 mm connectors 1250-1862
Connector cleaning
all connectors 92193Z
1 OOR 20 watt resistor capacitor removal
1. Order from Mountz Company, 1080 North 11th Street, San Jose, CA 95112.
2. Contains male and female gage and calibration block. These gages come with
mm HP 8510calibration kit.
If you own a calibration kit, refer to
kit manual for information on gaging and cleaning the connector.
3. Refer to the
Microwave
Connector
Care
Manual for cleaning instructions. Note, early versions of this manual
recommend cotton swabs, however, the new foam swabs provided in the connector cleaning kit should always
be used instead.
5-8 Test
Troubleshooting HP
Test