4. See Figure 3-8. Hold the cable with one hand and with the other, press the clamp gently and steadily
while wiggling the clamp straight over the RF connector and onto the test port connector nut. The
internal flats in the clamp must fit over the flats on the test port connector nut. Try not to rotate the
clamp as you do this or the RF connection may be loosened.
Figure
3-8.
Mating the Clomp and Nut Flats
5. See Figure 3-9.
Make sure that the thumbscrew is aligned with the counter-sunk hole in the clamp
body. Push the clamp toward the test set front panel and then tighten the thumbscrew with your
fingers. The cable cannot be damaged by tightening the thumbscrew too tightly.
CORRECT
INCORRECT
Figure 3-9.
Aligning the Thumbscrew With the Counter-bored Hole
6.
Repeat steps 1 through 5 for the other clamp.
This completes the anti-rotation clamp installation procedure. Refer to the installation section of the
test set operating and service manual for instructions to replace the internal O-ring in the anti-rotation
clamp.