Section 9. Adiustments
The HP8514B and
have one adjustment only, the sampler assembly adjustment. The
procedure is documented in the
Service Manual in the section titled Adjustments. It is also in
the Test Set Adjustments
section in volume 2 of the HP 8510A manual set. The adjustment is
guided.
This adjustment should be performed only if the need to do so is firmly established. A poor adjustment
is worse than none at all. Thus the software driven adjustment procedure begins with a check of the
sampler in question to establish the need.
Do not perform the actual adjustment if the sampler passes the check.
HP
Test Sets
Adjustments