64
SECTION 15 n.
CALIBRATE BASE UNIT CONT.
Documenting Calibration Data Procedure
NOTE: The input jacks for the 10 and 30 volt inputs are the
same input jacks, the accuracy statement for 10 volts is for
values between 0 and 10 volts DC, the accuracy statement for
30 volts DC is for values above 10 volts to 30 volts DC
VOLTAGE JACK INPUT, 0-10 volts
1. Connect the negative lead of the voltage standard to the
black ringed “COM” input jack
2. Connect the positive lead of the voltage standard to the red
ringed “0-33Vdc In” jack
3. Power on a voltage standard capable of delivering up to
10.000v DC and allow it to warm up to its specified accu-
racy as defined by the manufacturer
4. Power on the HHC using either USB or battery power and
allow it warm up to its specified accuracy (5 minutes)
5. Set the voltage standard to 0.000 volts dc
6. With the HHC in measure mode press the Ma/V key repeat-
edly until a voltage reading appears on the display
7. Make note of the voltage standard input voltage and the
voltage value displayed on the HHC display.
8. Verify that displayed voltage is within ±2.0 mV (±0.0020
volt)
9. I ncrease the input voltage setting from the voltage standard
at desired intervals until reaching 10.000 volts.
10. Note input standard voltages and HHC displayed voltages
and calculate error.
11. It is recommended to at minimum verify display voltage at
0.000 and 10.000 volts input.
12. When test is complete, set voltage standard to 0.00 volts
and disconnect the leads from the HHC.
VOLTAGE JACK INPUT, 10-30 volts
1. Connect the negative lead of the voltage standard to the
black ringed “COM” input jack
2. Connect the positive lead of the voltage standard to the red
ringed “0-33Vdc In” jack
3. Power on a voltage standard capable of delivering up to
30.000v DC and allow it to warm up to its specified accu-
racy as defined by the manufacturer
4. Power on the HHC using either USB or battery power and
allow it warm up to its specified accuracy (5 minutes)
5. Set the voltage standard to 10.000 volts dc
6. With the HHC in measure mode press the mA/V key
repeatedly until a voltage reading appears on the display
7. Make note of the voltage standard input voltage and the
voltage value displayed on the HHC display.
8. Verify that displayed voltage is within ±30 mV (±0.03 volt)
9. Increase the input voltage setting from the voltage standard
at desired intervals until reaching 30.000 volts.
10. Note input standard voltages and HHC displayed voltages
and calculate error.
11. It is recommended to at minimum verify display voltage at
10.00 and 30.00 volts input.