44
If the test is to be repeated press ZERO to immediately begin
another test.
To exit Leak Decay Teat mode press either ESC or MENU to
return to measurement mode.
The HHC can be used with most standard RTD (Resistive
Temperature Detector) probes to provide precision tempera-
ture measurements. An XXX –RT1 or XXX –RT2 Quick-Select
module is needed to interface an RTD probe to the HHC. There
are two XXX -RT Quick-Select module types available to sup-
port RTD resistance levels and capabilities indicated in the
table below:
Both modules can display temperature units of Fahrenheit,
Celsius, Kelvin, Rankine, and also support reading output in
units of ohms.
The electrical interface of the probe to the module incorporates
a Switchcraft TA4M type connector and requires an RTD with
the appropriate mating connector. The connector is available
through the supplier of your choice.
Setting up the PTE-2 for an RTD Temperature Measurement
Insert the appropriate XXX-RT Quick-Select RTD Interface
Module.
Connect the desired RTD probe to the Switchcraft connector
on the XXX-RT module.
Note: RTD probes can be connected using two, three or four
wire configurations. Use of four wire configuration is recom-
mended for optimum accuracy. RTD probes provided by this
manufacturer are of the four wire configuration.
XXX-RT1
XXX-RT2
Ohms Measurement
Range
0 - 400 ohms
0 - 4000 ohms
Factory Programmed
RTD Probe 01
Pt100 385: -200 to
500C ±0.15C
Pt1000: 184 to 275C
±0.15C
Factory Programmed
RTD Probe 02
Pt100 392: 550 to
850C ±0.2C
----
Factory Programmed
RTD Probe 03
Ni 120: 80 to 260C
± 0.1C
----
Factory Programmed
RTD Probe 04
Cu 10: 70 to 150C
± 0.6C
----
Available User
Programmable RTD
Probes
4
7
Ohms
+0.01% reading
+0.02ohms
+0.01% reading
+0.2ohms
SECTION 15 j.
RTD TEMPERATURE
MEASUREMENT
SECTION 15 i.
DECAY LEAK TEST CONT.
Summary of Contents for PTE-2
Page 2: ...ii...
Page 52: ...50 SECTION 15 l FIRMWARE UPDATE CONT...
Page 53: ...51 SECTION 15 l FIRMWARE UPDATE CONT...
Page 57: ...55 SECTION 15 l FIRMWARE UPDATE CONT Note The File Transfer can take up to 15 Minutes...
Page 86: ...84 NOTES...
Page 87: ...85...