22
SECTION 14 j.
WIRING DIAGRAMS
+
–
+
–
24Vdc Out
0-33Vdc In
0-22mA In
COM
24Vdc Out
0-33Vdc In
0-22mA In
COM
24Vdc Out
0-33Vdc In
0-22mA In
COM
RESISTOR*
EXTERNAL
24 VOLT DC
SWITCH HOOKUP
USING INTERNAL
SUPPLY SWITCH
NORMALLY OPEN
SWITCH HOOKUP
USING INTERNAL
SUPPLY SWITCH
NORMALLY OPEN
EXTERNAL
24 VDC SUPPLY
Note that you can use a normally closed switch as well
*Resistor should be sized to allow current as follows:
2 mA < I < 36 mA, 2 mA
12 k
, 35 mA
686
SWITCH HOOKUP
USING INTERNAL
SUPPLY SWITCH
NORMALLY OPEN
4/20 TRANSMITTER HOOK UP USING,
INTERNAL SUPPLY
4/20 TRANSMITTER HOOK UP USING,
EXTERNAL SUPPLY
4/20
TRANSMITTER
24 Vdc
EXTERNAL
SUPPLY
+ –
4/20ma
TRANSMITTER
+ –
+
–
+24V ma Comm Vdc
+24V ma Comm Vdc
V OUT TRANSDUCER HOOK UP USING,
INTERNAL SUPPLY
V OUT TRANSDUCER HOOK UP USING,
EXTERNAL SUPPLY
V OUT
TRANSDUCER
24 Vdc
EXTERNAL
SUPPLY
+ –
V OUT
TRANSDUCER
–
+
–
+24V ma Comm Vdc
+24 ma Comm Vdc
+
SECTION 14 i.
mA/V FUNCTION
The HHC has the capability to monitor either DC milliamp
current 0-22 mA or DC voltage 0-33 Vdc. Use of this feature
requires connection to the voltage or current source via the
4mm “Banana Jacks” located above the units LCD. Note that
this calibrator can monitor only one parameter, either voltage or
current, at a time.
To monitor voltage connect the negative lead of the voltage
source to the socket labeled “COM” with a black ring surround-
ing the socket. Connect the positive lead of the voltage source
to the red ringed terminal labeled “0-33Vdc In”.
To monitor current connect the negative lead to the black
ringed terminal labeled “COM. Connect the positive current
source to the red ringed terminal labeled “0-22mA In”.
Devices under test may be powered by an external power
source, or by the 24 Vdc power supply built into standard (non-
intrinsically safe) HHC base units. See wiring diagrams in sec-
tion 14j.
Press the mA/V key to activate voltage and current monitoring.
Press again to rotate through Volts, mA and pressure only
Summary of Contents for PTE-2
Page 2: ...ii...
Page 52: ...50 SECTION 15 l FIRMWARE UPDATE CONT...
Page 53: ...51 SECTION 15 l FIRMWARE UPDATE CONT...
Page 57: ...55 SECTION 15 l FIRMWARE UPDATE CONT Note The File Transfer can take up to 15 Minutes...
Page 86: ...84 NOTES...
Page 87: ...85...