Change direction: Hei-FLOW Advantage, Hei-FLOW Precision
✓
Left control light of direction key is
illuminated.
✓
Rotation is switched off.
➜
Press direction key.
✓
Right control light of direction key
illuminates.
✓
Pump will operate clockwise.
➜
Press direction key.
✓
Left control light of direction key
illuminates.
✓
Pump will operate counter-clockwise.
Clockwise
Counter-clockwise
Additional functions Hei-FLOW Precision
On top of the standard functions you can set parameters that allow consistent process
quality.
Following parameters can be set:
▪
„Speed rpm”
sets speed of pumping;
flow per minute “Flow ml/min” will be adjusted automatically
▪
„Tube mm” (inner diameter)
allows minimum and maximum flow per minute
must be set correctly for any other parameter to be adjusted correctly
▪
„Flow ml/min”
sets the flow per minute;
speed “Speed rpm” will be adjusted automatically
▪
„Vol ml”
sets the volume to be pumped in total;
pump will be stopped automatically when volume is reached
▪
„Pause sec/min/h”
intermittent dosing allows a permanent repetition of dispensing a defined volume
Correct operation of all functions requires the correct setting of pump
head and inner tube diameter in use.
When purchasing a package with pump head, the pump head is set
correctly by the factory.
Set pump head
Factory pre-setting is SP quick (Pu 1).
Display Pump head
Hei-FLOW
Precision 01
Hei-FLOW
Precision 06
Pu 0
Multi-channel
C 8
✓
-
Pu 1
Single-channel
SP quick
✓
✓
Pu 2
Single-channel
SP standard / SP vario
✓
✓
Pu 3
not assigned
-
✓
-
Pu 4
not assigned
-
✓
-
Pu 8
Multi-channel
C 12
✓
-
Pu 9
Multi-channel
C 4
✓
-
➜
Switch unit off at main switch.
➜
Press „Cal” key and hold pressed.
➜
Switch unit on at main switch.
✓
Display shows actual setting of pump
head.
➜
Release „Cal” key.
➜
Select required pump head via arrow
keys or .
➜
Confirm selection by pressing „Select”
key.
Set tube
Possible inner diameters of tubes:
(also see chapter „Technical data, Selection table tubes”)
0.8
1.7
3.1
4.8
6.4
7.9
✓
Unit is switched on.
➜
Press „Select” key repeatedly, till
indicator light „Tube mm” illuminates.
➜
Select required inner diameter of tube
in use via arrow keys or .
➜
Confirm selection by pressing „Select”
key.
A blinking indicator light „Vol ml” during setting of tube diameter
shows that the device adjusts volume to required minimum of new
diameter.