Assembly
Assembly
28
29
Assembly
Electrical connections
▪
Repairs
of any kind are allowed by
qualified professional electricians only
.
Any improper repair can result in a dangerous situation. Contact your local Heidolph
Instruments distributor for any repair you may have.
Warning:
Electric shock!
This unit must only be connected to a grounded electrical socket.
The unit and voltage must match. The type label on the back side of
the unit provides all voltage specifics.
When shipped the unit is grounded. In case the original plug is
changed, the new plug must have a protective conductor!
Rating plate
ON/OFF switch
Color code for electrical connections:
Europe
North
America
GREEN/
YELLOW
PE: Protective
conductor
(Earth)
GREEN
PE: Protective
conductor
(Earth)
BLUE
N: Neutral conductor
WHITE
N: Neutral conductor
1,411 mm
P: Phase
BLACK
P: Phase
The unit is connected with the power cord supplied to the electrical socket. It plugs into
the housing on the rear side of the unit.
For countries utilizing a plug other than the standard one supplied:
▪
The plug supplied may only be changed by a professional electrician.
▪
If utilizing an adaptor ensure that it is grounded and approved by the local regula-
tions.
Mount pump heads
Warning:
Risk of finger crushing!
When mounting pump heads you can crush your fingers if rotation is
switched on accidentally.
Always switch device off via main switch while mounting pump
heads.
Single-channel pump heads SP standard and SP vario
Components:
▪
1x Pump head
▪
1x Flange
▪
1x Guide aid
▪
2x Knurled screws
▪
2x Short tube sections
▪
1x Allen key
Pump head
Flange
Knurled screws
Allen key
Guide aids
Tube
sections
➜
Insert guide aid into flange.
➜
Position flange with guide aid in front
of pump head in a way that guide will
be pushed into coupling.
✓
Allen screw shows upwards.
➜
Tighten flange on both sides with
knurled screws onto pump head.
➜
Remove guide aid.
Knurled screws
Allen screw
Guide aids
Coupling
➜
Remove protection shield from front of
pump head.
Warning:
Risk of finger crushing!
When pump head is not
covered, you can crush your
fingers on rotating parts.
Always mount protection shield
before switching on and
operating device.
Protection shield
Knurled screws
Summary of Contents for Hei-FLOW Advantage 01