Attachments
56
Cooling unit
Danger!: Risk of injury! Risk of poisoning! Environmental haz-
ard!
The container of the cooling unit is under considerable pressure (19-28
bar) and contains environmentally harmful substances. Improper
emptying of the refrigerant could result in a serious risk of injury or
poisoning as well as environmental hazards.
The refrigerant in the internal cooling circuit may only be emptied by a
qualified refrigeration technician.
The compressor in the cooling unit is hermetically sealed and equipped with overload
protection that responds to the temperature and power consumption of the compressor.
The cooling unit contains refrigerant. Observe the special regulations
for its disposal (see also chapter “Safety, Disposal”).
Cooling output in kW at ambient temperatures in °C
Cooling output kW
Hei-CHILL 3000
Hei-CHILL 5000
at 20 °C
3,00
5,00
at 10 °C
2,40
3,90
at 0 °C
1,68
2,75
at -10 °C
1,03
1,70
at -20 °C
0,60
1,00
* The cooling output is measured with the thermal fluid maintained at a certain temperature. The ambient
temperature for the measurements is 20 °C.
Refrigerant and fill level
The cooling unit contains refrigerant. Observe the special regulations
for its disposal (see chapter “Safety, Disposal”).
The refrigerant R-449A
contains fluorinated greenhouse gases.
The global warming potential (GWP) is 1430 at a time horizon of 100
years, CO2
equivalent = 1.0.