Operation
30
Adjusting the internal control parameters
✓
The device is switched on.
➜
Select the submenu “Setup” /
“Control” / “Control parameters” /
”Internal Pt1000”.
➜
Use the arrow buttons to select one
of the presented options and confirm
your selection using the button
“Select” .
✓
By selecting "Tv manual/auto" you
can specify the option of automatic or
manual control.
✓
When selecting any other option, you
will be able to adjust the highlighted
values.
✓
Confirm the changed value using the
button “Select” .
If the control parameters “Tv manual/auto” is set to automatic control,
the subsequent parameters “Tv” and “Td” are derived using factors
from Tn. They are blocked from changes and indicated by a lock
symbol.