Start-Up
15
Filling for the first time
The reservoir for the thermal fluid can hold between 20 and 33 litres (see chapter
“Attachments, Technical data”).
Filling of the device is supported by a software program. If the fill level is too low:
▪
the pump and cooling unit will not start
▪
the Fill mode starts automatically.
➜
Close the drain valve by turning the valve
fully to the left.
Emptying cock closed
✓
The device is switched on.
✓
The " Fill mode" submenu appears on
the display.
➜
Carefully remove the cover for the
filling pipe by pulling it upwards.
➜
Carefully fill the reservoir with thermal
fluid via the filling pipe using a funnel.
Sealing cap for reservoir
Filling pipe for reservoir
✓
The display shows the value "0" as the
fill level to the right of "started".
➜
Observe the fill level indicator.
➜
An acoustic signal sounds at intervals
once the value has risen to 4-5.
✓
The device is filled sufficiently.
➜
Connect your external consumer (see chapter “Assembly, Peripheral connections”).
➜
Press the "Standby" softkey to fill the external consumer.
➜
Continue to refill thermal fluid until the fill level rises to 4-5 again and an acoustic signal
will sound.
➜
End the filling process and close the reservoir with the cover.
➜
Press the down
ˇ
arrow button to select “Exit filling”.
➜
Press the "ESC" softkey to exit the "Fill mode" menu.
✓
You return to the base menu.
The intervals between the signal tones are shortened until a contin-
uous tone is heard if the filling process does not end. A continuous tone
indicates that the device is overfilled and cannot be started. Before
starting the pump, some thermal fluid must be drained first
Summary of Contents for Hei-CHILL 3000