Connecting QuantumX modules
CX22B / CX22B-W
A3169-8.0
HBM: public
27
5
Connecting QuantumX modules
The simplest way to connect QuantumX modules to the
Data Recorder is via FireWire. Connect cable
1-KAB272-x to any of the Data Recorder's connections
and the other end to connection X102 of the nearest
measurement module.
Then keep threading: connection X102 to X101 of the
second measurement module, etc. Remember the maxi
mum current of 1.5 A. As a rule, you must introduce a
voltage supply again after 3 modules
.
You can also con
nect measurement modules at the other connection.
Alternatively, you can connect all the QuantumX modules
to the CX22B‒W via Ethernet. If you have more than two
modules, you will also require an Ethernet switch.
In the configuration shown below, the module supply volt
age is looped through via FireWire (max. 1.5 A via
FireWire). The power consumption of a module is ex
plained in the data sheet
.
You can connect up to 12 modules to the CX22B‒W
Data Recorder via FireWire, with a supply feed required
for at least every 4th module, with approximately the
same voltage value. The general QuantumX operating
manual contains more detailed data on this
.