Configuring the CX22B‒W for self-contained measurements
CX22B / CX22B-W
A3169-8.0
HBM: public
65
► Exit the dialog by clicking on
OK
.
► Exit catman
EASY.
► In the start screen, click on
Settings
.
► On the
General
tab, activate the option
Launch DAQ
software (catmanEASY) at CX22B‒W startup
and
click on
OK
.
► Click on
Commit changes and reboot
to perma
nently save the new settings.
This project will then start automatically the next time the
CX22B‒W is switched on. If the power supply is inter
rupted during measurement, the project will be reloaded
and measurement continued when the power supply is
restored. The module restart, up to the continuation of
measurement, will require approx. 3 minutes, depending
on the number of connected devices and channels.
Important
The data already present in the catman
[
EASY tempo
rary measurement data memory are not lost if the power
supply is
interrupted. Any additional files will be assigned
a counter at the end of the file name.