Safety instructions
ITC-K1000
A4749-1.0
HBM: public
11
tection provided by the thermocouple outside of the mea
surement categories defined in EN 61010 is not
guaranteed above these voltages.
DANGER
In the measurement category II range, the ITC-K100 is
approved for a maximum voltage to ground potential of
600 V DC or 600 V AC rms.
The protection provided by the thermocouple within this
measurement category is not guaranteed above these
voltages.
DANGER
In the measurement category III range, the ITC-K1000 is
approved for a maximum voltage to ground potential of
300 V DC or 300 V AC rms.
The protection provided by the thermocouple within this
measurement category is not guaranteed above these
voltages.
DANGER
Operation of the ITC-K1000 thermocouple in the range of
measurement category IV is not permitted.
The protection provided by the thermocouple is not
guaranteed within measurement category IV.
DANGER
The protection provided by the ITC-K1000 thermocouple
can be impaired if the thermocouple is operated in
ambient air that is too humid or has too much condensa
tion, or in damp or wet conditions, such as after liquid
spillages (see specifications).
Summary of Contents for Quantum X MX403B
Page 7: ...MX403B A3757 2 2 HBM public 5...
Page 38: ...36 A3757 2 2 HBM public MX403B...
Page 76: ...38 A3757 2 2 HBM public MX403B...
Page 283: ...Caract ristiques techniques MX809B A3757 2 2 HBM public 57...
Page 285: ...Operating Manual Bedienungsanleitung English Deutsch ITC U1001...
Page 306: ...Specifications 20 A4749 1 0 HBM public ITC U1001...
Page 327: ...Technische Daten ITC U1001 A4749 1 0 HBM public 21...
Page 350: ...Caract ristiques techniques 22 A4749 1 0 HBM public ITC U1001...
Page 351: ...Caract ristiques techniques ITC U1001 A4749 1 0 HBM public 23...
Page 374: ...Specifications 20 A4749 1 0 HBM public ITC K1000...
Page 415: ...Caract ristiques techniques ITC K1000 A4749 1 0 HBM public 21...