28
A3757-2.2
HBM: public
MX403B
6.2
Connection/measurement
► Before connecting the voltage source to be measured,
check that the MX403B is working. This is done with a
calibrated reference source or by measuring the refer
ence voltage in advance with a calibrated voltmeter.
► Connect the MX403B to the
voltage source to be
measured
.
- Only extend the measuring lead with measuring
leads approved for CAT II 1000 V or CAT III 600 V,
or better.
- The maximum permissible length of the measuring
lead is 30 m.
- The measuring lead must not run through outdoor
areas, either in full or in part.
► To reduce interference signals:
- Keep the measuring leads as short as possible.
- Connect the MX403B housing to ground potential.
- Use pairs of shielded cables (ensure approval),
connect the shield to ground potential.
- Route unshielded measuring leads close together,
channel by channel; avoid large loop areas.
- Twist the unshielded measuring leads together.
- Physically separate leads for measuring sensitive,
small signals from measuring leads connected to
sources that are subject to interference or highly
dynamic.
- Do not route measuring leads next to potentially
interfering devices such as motors, inverters, etc.
Summary of Contents for Quantum X MX403B
Page 7: ...MX403B A3757 2 2 HBM public 5...
Page 38: ...36 A3757 2 2 HBM public MX403B...
Page 76: ...38 A3757 2 2 HBM public MX403B...
Page 283: ...Caract ristiques techniques MX809B A3757 2 2 HBM public 57...
Page 285: ...Operating Manual Bedienungsanleitung English Deutsch ITC U1001...
Page 306: ...Specifications 20 A4749 1 0 HBM public ITC U1001...
Page 327: ...Technische Daten ITC U1001 A4749 1 0 HBM public 21...
Page 350: ...Caract ristiques techniques 22 A4749 1 0 HBM public ITC U1001...
Page 351: ...Caract ristiques techniques ITC U1001 A4749 1 0 HBM public 23...
Page 374: ...Specifications 20 A4749 1 0 HBM public ITC K1000...
Page 415: ...Caract ristiques techniques ITC K1000 A4749 1 0 HBM public 21...