Specifications
54
A3757-2.2
HBM: public
MX809B
Electric voltage
"
5 V
Accuracy class
0.02
Transducers that can be con
nected
Voltage sources up to
"
5 V
Measurement range
V
"
5
Permissible Input voltage
V
"
15
Permissible cable length
between
MX809B and measurement location,
max.
m
30
Measurement frequency range (-3
dB)
Hz
0 ... 55
Internal resistance of voltage
source
W
< 500
Input impedance, typical
M
W
> 2.5
Noise at 25
5
C (peak-to-peak)
with 1 Hz Bessel filter
with 10 Hz Bessel filter
with filter OFF, 1000 S/s
mV
mV
mV
< 0.1
< 0.2
< 0.3
Non-linearity
%
< 0.02 of full scale value
Common-mode rejection
for UCM_RMS 707V, 80 Hz
dB
> 100
Zero drift
% /
10 K
< 0.01 of full scale value
Full-scale drift
% /
10 K
< 0.02 of measured value