Internal calculation channels
188
A4354-2.2
HBM: public
PMX
13.2.2.3 Extreme Value (Peak)
Function
This calculation determines the minimum, maximum or
oscillation width (peak-to-peak) value of a signal. You can also
determine the value of another channel (Input 2) when an
extreme value is reached. If you specify a discharge time
constant, you will obtain an envelope function.
Additionally, a new peak is signalled by an output flag. The
flag only is high in the cycle, in which a new peak was
detected. When the input signal increases constantly, the
output flag of a maximum peak block is permanently high.
Reset via (virtual) digital input or Flag (levelcontrolled)
Reset via fieldbus (eventcontrolled)
Hold via digital input (levelcontrolled)
This function block may deliver wrong values after its creation
or after switching of the parameter set. It is strongly
recommended to reset the extreme value before starting the
process.
The maximum or peaktopeak value decreases with the
discharge rate.
The minimum value increases with the discharge rate.
Small discharge rates can cause numerical issues. Please
compare the discharge rate with the expected peak values.
Discharge rate > 5% of peak value: No issue expected.
Discharge rate 1% ... 5% of peak value: Please check whether
the value discharges as expected.
Discharge rate < 1% of peak: Critical. The value may not
discharge as expected.
Inputs
- Input 1
- Input 2
-
Hold with
- Reset by
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...