Hameg HM304 Manual Download Page 23

23

Subject to change without notice

Test Instructions

General

These Test Instructions are intended as an aid for checking
the most important characteristics of the HM304 at regular
intervals without the need for expensive test equipment.
Resulting corrections and readjustments inside the
instrument, indicated by the following tests, are described in
the Service Instructions or on the Adjusting Plan. They should
only be undertaken by qualified personnel.

As with the First Time Operation instructions, care should be
taken that all knobs with arrows are set to their calibrated
positions. Depress AUTO SET for default settings. It is
recommended to switch on the instrument for about 20
minutes prior to the commencement of any check.

Cathode-Ray Tube: Brightness and Focus,
Linearity, Raster Distortions

Normally, the CRT of the HM304 has very good brightness.
Any reduction of this brightness can only be judged visually.
However, decreased brightness may be the result of wrong
settings or reduced high voltage. The latter is easily
recognized by the greatly increased sensitivity of the vertical
amplifier. Correct setting means, that the HOLD OFF control
should be turned to the left stop; the X-MAG. x10 function
should be switched off; a medium time coefficient should be
selected; line triggering (~ indicated) should be used only
with a suitable TIME/DIV. setting (e.g. 2ms/div.). The control
range for maximum and minimum brightness (intensity) must
be such that the beam just disappears before reaching the
left hand stop of the INTENS. control (particularly when in
XY mode), while with the control at the right hand stop the
focus and the line width are just acceptable.

With maximum intensity the timebase fly-back must on no
account be visible. Visible trace fault without input signal:
bright dot on the left side or decreasing brightness from left
to right or shortening of the baseline. (Cause: Incorrect
Unblanking Pulse.)
It should be noted that with wide variations in brightness,
refocusing is always necessary. Moreover, with maximum
brightness, no „pumping“ of the display must occur. If
pumping does occur, it is normally due to a fault in the
regulation circuitry for the high voltage supply. The presetting
pots for the high voltage circuit, minimum and maximum
intensity, are only accessible inside the instrument.

A certain out-of-focus condition in the edge zone of the screen
must be accepted. It is limited by standards of the CRT
manufacturer. The same is valid for tolerances of the
orthogonality, the undeflected spot position, the non-linearity
and the raster distortion in the marginal zone of the screen in
accordance with international standards (see CRT data book).
These limit values are strictly supervised by HAMEG. The
selection of a cathode-ray tube without any tolerances is
practically impossible.

Astigmatism Check

Check whether the horizontal and vertical sharpness of the
display are equal. This is best seen by displaying a square-
wave signal with the repetition rate of approximately 1MHz.
Focus the horizontal tops of the square-wave signal at nor-
mal intensity, then check the sharpness of the vertical edges.
If it is possible to improve this vertical sharpness by turning
the FOCUS control, then an adjustment of the astigmatism
control is necessary. A potentiometer of 47k

 is provided

inside the instrument for the correction of astigmatism. A
certain loss of marginal sharpness of the CRT is unavoidable;
this is due to the manufacturing process of the CRT.

Symmetry and Drift of the Vertical Amplifier

Both of these characteristics are substantially determined
by the input stages of the amplifiers.

The symmetry of both channels and the vertical final amplifier
can be checked by inverting Channel I and II (depress the
corresponding INV pushbutton). The vertical position of the
trace should not change by more than 0.5div. However, a
change of 1div. is just permissible. Larger deviations indicate
that changes have occurred in the amplifier.

A further check of the vertical amplifier symmetry is possible
by checking the control range of the Y-POS. controls. A sine-
wave signal of 10-100kHz is applied to the amplifier input. When
the Y-POS. control is then turned fully in both directions from
stop to stop with a display height of approximately 8div., the
upper and lower positions of the trace that are visible should
be approximately of the same height. Differences of up to
1div. are permissible (input coupling should be set to AC).

Checking the drift is relatively simple. 20minutes after
switching on the instrument, set the baseline exactly on the
horizontal center line of the graticule. The beam position must
not change by more than 0.5div. during the following hour.

Calibration of the Vertical Amplifier

Two square-wave voltages of 0.2Vpp ±1% and 2Vpp are
present at the output sockets of the calibrator (CAL.) If a
direct connection is made between the 0.2V output and the
input of the vertical amplifier (e.g. using a x1 probe), the
displayed signal in the 50mV/div. position (variable control to
CAL.) should be 4div. high (DC input coupling). Maximum
deviations of 0.12div. (3%) are permissible. If a x10 probe  is
connected between the 2V output and Y input, the same
display height should result.
With higher tolerances it should first be investigated whether
the cause lies, within the amplifier or in the amplitude of the
square-wave signal. On occasions it is possible that the pro-
be is faulty or incorrectly compensated. If necessary the
measuring amplifier can be calibrated with an accurately
known DC voltage (DC input coupling). The trace position
should then vary in accordance with the deflection coefficient
set.

With variable control in the attenuator sector fully counter-
clockwise, the input sensitivity is decreased at least by the
factor 2.5 in each position. In the 50mV/div. position, the
displayed calibrator signal height should vary from 4div. to at
least 1.6div.

Transmission Performance of the
Vertical Amplifier

The transient response and the delay distortion correction
can only be checked with the aid of a square-wave generator
with a fast risetime (max. 5ns). The signal coaxial cable (e.g.
HZ34) must be terminated at the vertical input of the
oscilloscope with a resistor equal to the characteristic
impedance of the cable (e.g. with HZ22). Checks should be
made at 100Hz, 1kHz, 10kHz, 100kHz and 1MHz, the
deflection coefficient should be set at 5mV/div. with DC input
coupling (Y variable control in CAL. position). In so doing, the
square pulses must have a flat top without ramp-off, spikes
and glitches; no overshoot is permitted, especially at 1MHz
and a display height of 4-5div.. At the same time, the leading
top corner of the pulse must not be rounded. In general, no
great changes occur after the instrument has left the factory,
and it is left to the operators discretion whether this test is
undertaken or not. A suited generator for this test is HZ60
from HAMEG.

Summary of Contents for HM304

Page 1: ...tructions General Information 5 Symbols 5 Use of tilt handle 5 Safety 5 Operating conditions 5 EMC 6 Warranty 6 Maintenance 6 Protective Switch Off 6 Power supply 6 Type of signal voltage 7 Amplitude...

Page 2: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection...

Page 3: ...nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized...

Page 4: ...35MHz 3dB Risetime 10ns Overshoot max 1 Deflection coefficients 14 calibrated steps from1mV div to20V div 1 2 5 sequence with variable 2 5 1 up to 50V div Accuracy in calibrated position 1mV div to 2m...

Page 5: ...three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact T...

Page 6: ...aintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largely certain that all signals are displayed with the accuracy on...

Page 7: ...ms Veff have 2 83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective va...

Page 8: ...e base setting indicated by one of the TIME DIV LED s one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div when the red sec LED and the 0...

Page 9: ...hes next to the edge are also not taken into account With very severe transient distortions the rise and fall time measurement has little meaning For amplifiers with approximately constant group delay...

Page 10: ...oscilloscope by depressing the red POWER pushbutton The instrument will revert to its last used operating mode Except in the case of COMP TESTER mode where a trace appears on the screen if the INTENS...

Page 11: ...should be adjusted The location of the low frequency compensation trimmer can be found in the probe information sheet Adjust the trimmer with the insulated screw driver provided until the tops of the...

Page 12: ...channell I to channel II and vice versa after each sweep period In DUAL mode the internal trigger source can be switched over from channel I to channel II and vice versa if the TRIG pushbutton is dep...

Page 13: ...compared with C short circuited then the test voltage leads the reference voltage and vice versa This applies only in the region up to 90 phase shift Therefore C should be sufficiently large and produ...

Page 14: ...t when external trigger is used the trigger threshold can be stated as vertical display height in div through which the time base generator starts the display is stable and the trigger LED located in...

Page 15: ...gnals than the DC coupling because the white noise in the trigger voltage is strongly suppressed So jitter or double triggering of complex signals is avoidable or at least reduced in particular with v...

Page 16: ...er voltage may have a completely different form from the test signal voltage Triggering is even possible in certain limits with whole number multiples or fractions of the test frequency but only with...

Page 17: ...following explanation assumes that the trace starts on the left vertical graticule line Photo 1 composite video signal MODE undelayed TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Depre...

Page 18: ...ted in Dual mode under conditions where DUAL chopped mode is active this display mode is not switched off when time coefficients are being reduced 0 2ms div to 0 05 s div for signal expansion in DEL a...

Page 19: ...also be made to integrated circuits All these components can be tested in and out of circuit The test principle is fascinatingly simple A built in generator delivers a sine voltage which is applied a...

Page 20: ...ore testing of transistor amplification is not possible but testing of a single junction is easily and quickly possible Since the test voltage applied is only very low all sections of most semiconduct...

Page 21: ...ould then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational...

Page 22: ...Subject to change without notice 22...

Page 23: ...of astigmatism A certain loss of marginal sharpness of the CRT is unavoidable this is due to the manufacturing process of the CRT Symmetry and Drift of the Vertical Amplifier Both of these characteri...

Page 24: ...eck of the mono channel display is unnecessary it is contained indirectly in the tests above stated Triggering Checks The internal trigger threshold is important as it determines the display height fr...

Page 25: ...and also the influence of the earths magnetic field which is dependent on the instruments North orientation are corrected by means of the TR potentiometer In general the trace rotation range is asymme...

Page 26: ...hanged Generally max halving or doubling of this resistance value should be sufficient A too small trigger threshold cause double triggering or premature trigger action due to interference pulses or r...

Page 27: ...R ckgabe Beschreibung PC Scope Scope PC ID ID Daten CR LF data consits of instrument type manufacturer CR R CR LF remote status and baud rate acceptance TRSTA TRSTA b CR LF query for trigger status d...

Page 28: ...nter 0 13 CH2 GND AC INV2 ON VALUE Counter 0 13 mv DIV 0000 20V DIV 1101 MODE CT XY A TR CHOP ADD 0 TR SOURCE 00 Y1 01 Y2 1x EXT TB1 x10 0 0 TB A TIME Counter 1 26 50ns DIV 00 bis 0 5s DIV 15hex TB2 0...

Page 29: ...cy ranges AC 10Hz 100MHz DC 0Hz 100MHz HF 1 5kHz 100MHz LF 0Hz 1 5 kHz TV L to trigger on line sync pulses TV F to trigger on separated frame sync pulses Select SLOPE for the leading slope Sync pulse...

Page 30: ...control for trace sharpness mechanical knob TR Trace rotation mechanical To align trace with horzontal field potentiometer graticule line Compensates adjustment with influence of Earth s magnetic scr...

Page 31: ...triggering Not available in combination with ext triggering XY or COMP TESTER modes INPUT CH II Channel II signal input BNC connector Input impedance 1M II 20pF AC DC Selects input coupling of CH II...

Reviews: