Hameg HM304 Manual Download Page 12

Subject to change without notice

12

The adjustment sequence must be followed in the order
described, i.e. first at 1kHz, then at 1MHz. The calibrator
frequencies should not be used for timebase calibration. The
pulse duty cycle deviates from 1:1 ratio.

Prerequisites for precise and easy probe adjustments, as well
as checks of deflection coefficients, are straight horizontal
pulse tops, calibrated pulse amplitude, and zero-potential at
the pulse base. Frequency and duty cycle are relatively
uncritical. For interpretation of transient response, fast pulse
risetimes and low-impedance generator outputs are of
particular importance.

Providing these essential features, as well as switch-
selectable output-frequencies, the calibrator of the HM304
can, under certain conditions, replace expensive squarewave
generators when testing or compensating wideband-
attenuators or -amplifiers. In such a case, the input of an
appropriate circuit will be connected to one of the 

CAL.

-

outputs via a suitable probe.

The voltage provided at a high-impedance input (1M

 II 15-

30pF) will correspond to the division ratio of the probe used
(10:1 = 20mVpp, 100:1 = also 20mVpp from 2V output).
Suitable probes are HZ51, 52, 53, and 54.

Operating modes of the vertical
amplifiers in Yt mode.

The vertical amplifier is set to the desired operating mode by
using the 2 pushbuttons 

CH I

 and  

CH II

 (for 

CH I

CH II

,

DUAL

 and 

ADD

 mode) in the Y field of the front panel. The

different modes are indicated by LED‘s in the 

channel I

 and

channel II

 

VOLTS/DIV

 sections and the 

ADD LED

 in ADD

mode.
If only 

CH II

 is active to switch to CH I mode, first press the

CH I

 pushbutton to switch on channel I. Now the oscilloscope

is in 

DUAL

 mode where LED‘s in both 

VOLTS/DIV

 sectors

are lit. Then the CH II pushbutton must be depressed once
to switch off channel II. It is not possible to operate the
oscilloscope with both channels switched off. That is why
the required channel must first be switched on and then the
unwanted channel must be switched off.
To switch from CH I to CH II mode, first switch on CH II and
then switch off CH I. If  internal triggering is selected (EXT
LED near the 

TRIG. INP.

 socket extinguished), the trigger

source indicator LED‘s (

TR I

 and 

TR II

) will be switched over

simultaneously.

DUAL

 mode is selected if a LED is lit in each 

VOLTS/DIV

sectors. As mentioned before, one channel is always present
and so the other channel must be switched on for DUAL
mode operation.
In 

DUAL

 mode both channels are working. Two signals can

be displayed together in alternate or chopped mode. The
alternate mode is not suitable for displaying very slow-running
processes. The display then flickers or appears to jump.
Therefore the instrument automatically switches over from
alternate to chopped mode if 

TIME/DIV

 settings from 0.5ms/

div to 0.5s/div are used. If in chopped DUAL mode, both
channels are switched over constantly at a high frequency
within a sweep period. Low frequency signals below 1kHz,
or with periods longer than 1ms are then displayed without
flicker. Conversely in 

DUAL

 alternate mode, the displayed

channel switches over from channell I to channel II and vice
versa after each sweep period.
In DUAL mode the internal trigger source can be switched
over from channel I to channel II and vice versa if the 

TRIG.

pushbutton is depressed for a short time. Depressing the

TRIG.

 pushbutton in DUAL mode for a longer time switches

over to alternate triggering and consequently both TR I and
TR II LED‘s are lit. As alternate triggering is not possible in
combination with DUAL chopped mode, the instrument
automatically switches over to the alternate  mode if DUAL
chopped mode was active before. Alternate triggering can
be switched off by depressing the 

TRIG.

 pushbutton for a

short time. Then just one TR LED is lit.

DUAL 

chopped mode is also automatically switched off when

TV-F (television frame triggering) is selected to avoid
interference.
In combination with delay and triggered delay mode, DUAL
chopped mode can also be switched to DUAL alternate mode
by simultaniously depressing both pushbuttons marked with
< and > arrows in the TIME/DIV sector. Any change in the
delay mode time base setting reverts to the DUAL chopped
mode.

ADD mode is selected by simultaneously depressing both

CH I

 and 

CH II

 pushbuttons which causes the 

ADD LED

between both pushbuttons to light.
In 

ADD

 mode the signals of both channels are algebraically

added (±I  ±II) and displayed as one signal. Whether the
resulting display shows the sum  or difference is dependent
on the phase relationship or the polarity of the signals and on
the invert function indicated by the 

INV LED‘s

 for each

channel. To quit the 

ADD

 mode, depress the pushbutton for

the required channel or depress both CH I and CH II
pushbuttons for a short time to switch back to DUAL mode.
As alternate triggering is not available in ADD mode, the
instrument switches over from ADD mode to alternate DUAL
mode if the TRIG pushbutton is depressed for a longer time.

In 

ADD

 mode the following combinations are possible for

In-phase input voltages:

• Both INV (invert) function CH.I and

INV (invert) function CH.II active

• released or depressed = sum.
• Only one INV (invert) function active = difference.

Antiphase input voltages:

• Both INV (invert) function active

or inactive = difference.

• INV (invert) function CH.I  or  INV

(invert) function CH.II active = sum.

In the 

ADD

 mode the vertical display position is dependent

upon the 

Y-POS.

 setting of both channels. The same Y

deflection coefficient is normally used for both channels with
algebraic addition.

Please note that the Y-POS. settings are also added
but are not affected by the INV setting.

Differential measurement techniques allow direct
measurement of the voltage drop across floating components
(both ends above ground). Two identical probes should be
used for both vertical inputs. In order to avoid ground loops,
use a separate ground connection and do not use the probe
ground leads or cable shields.

X-Y Operation

For X-Y operation, the pushbutton in the X field marked XY
must be depressed. Then the XY LED in the 

TIME/DIV

 sector

is lit and the time coefficient indication is switched off. The X
signal is then derived from the 

INPUT CH II (X)

. The

calibration of the X signal during X-Y operation is determined
by the setting of the Channel II Y deflection coefficient and
variable control.
This means that the sensitivity ranges and input impedances

Summary of Contents for HM304

Page 1: ...tructions General Information 5 Symbols 5 Use of tilt handle 5 Safety 5 Operating conditions 5 EMC 6 Warranty 6 Maintenance 6 Protective Switch Off 6 Power supply 6 Type of signal voltage 7 Amplitude...

Page 2: ...uction in the manual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection...

Page 3: ...nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized...

Page 4: ...35MHz 3dB Risetime 10ns Overshoot max 1 Deflection coefficients 14 calibrated steps from1mV div to20V div 1 2 5 sequence with variable 2 5 1 up to 50V div Accuracy in calibrated position 1mV div to 2m...

Page 5: ...three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact T...

Page 6: ...aintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largely certain that all signals are displayed with the accuracy on...

Page 7: ...ms Veff have 2 83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective va...

Page 8: ...e base setting indicated by one of the TIME DIV LED s one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div when the red sec LED and the 0...

Page 9: ...hes next to the edge are also not taken into account With very severe transient distortions the rise and fall time measurement has little meaning For amplifiers with approximately constant group delay...

Page 10: ...oscilloscope by depressing the red POWER pushbutton The instrument will revert to its last used operating mode Except in the case of COMP TESTER mode where a trace appears on the screen if the INTENS...

Page 11: ...should be adjusted The location of the low frequency compensation trimmer can be found in the probe information sheet Adjust the trimmer with the insulated screw driver provided until the tops of the...

Page 12: ...channell I to channel II and vice versa after each sweep period In DUAL mode the internal trigger source can be switched over from channel I to channel II and vice versa if the TRIG pushbutton is dep...

Page 13: ...compared with C short circuited then the test voltage leads the reference voltage and vice versa This applies only in the region up to 90 phase shift Therefore C should be sufficiently large and produ...

Page 14: ...t when external trigger is used the trigger threshold can be stated as vertical display height in div through which the time base generator starts the display is stable and the trigger LED located in...

Page 15: ...gnals than the DC coupling because the white noise in the trigger voltage is strongly suppressed So jitter or double triggering of complex signals is avoidable or at least reduced in particular with v...

Page 16: ...er voltage may have a completely different form from the test signal voltage Triggering is even possible in certain limits with whole number multiples or fractions of the test frequency but only with...

Page 17: ...following explanation assumes that the trace starts on the left vertical graticule line Photo 1 composite video signal MODE undelayed TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Depre...

Page 18: ...ted in Dual mode under conditions where DUAL chopped mode is active this display mode is not switched off when time coefficients are being reduced 0 2ms div to 0 05 s div for signal expansion in DEL a...

Page 19: ...also be made to integrated circuits All these components can be tested in and out of circuit The test principle is fascinatingly simple A built in generator delivers a sine voltage which is applied a...

Page 20: ...ore testing of transistor amplification is not possible but testing of a single junction is easily and quickly possible Since the test voltage applied is only very low all sections of most semiconduct...

Page 21: ...ould then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which is known to be operational...

Page 22: ...Subject to change without notice 22...

Page 23: ...of astigmatism A certain loss of marginal sharpness of the CRT is unavoidable this is due to the manufacturing process of the CRT Symmetry and Drift of the Vertical Amplifier Both of these characteri...

Page 24: ...eck of the mono channel display is unnecessary it is contained indirectly in the tests above stated Triggering Checks The internal trigger threshold is important as it determines the display height fr...

Page 25: ...and also the influence of the earths magnetic field which is dependent on the instruments North orientation are corrected by means of the TR potentiometer In general the trace rotation range is asymme...

Page 26: ...hanged Generally max halving or doubling of this resistance value should be sufficient A too small trigger threshold cause double triggering or premature trigger action due to interference pulses or r...

Page 27: ...R ckgabe Beschreibung PC Scope Scope PC ID ID Daten CR LF data consits of instrument type manufacturer CR R CR LF remote status and baud rate acceptance TRSTA TRSTA b CR LF query for trigger status d...

Page 28: ...nter 0 13 CH2 GND AC INV2 ON VALUE Counter 0 13 mv DIV 0000 20V DIV 1101 MODE CT XY A TR CHOP ADD 0 TR SOURCE 00 Y1 01 Y2 1x EXT TB1 x10 0 0 TB A TIME Counter 1 26 50ns DIV 00 bis 0 5s DIV 15hex TB2 0...

Page 29: ...cy ranges AC 10Hz 100MHz DC 0Hz 100MHz HF 1 5kHz 100MHz LF 0Hz 1 5 kHz TV L to trigger on line sync pulses TV F to trigger on separated frame sync pulses Select SLOPE for the leading slope Sync pulse...

Page 30: ...control for trace sharpness mechanical knob TR Trace rotation mechanical To align trace with horzontal field potentiometer graticule line Compensates adjustment with influence of Earth s magnetic scr...

Page 31: ...triggering Not available in combination with ext triggering XY or COMP TESTER modes INPUT CH II Channel II signal input BNC connector Input impedance 1M II 20pF AC DC Selects input coupling of CH II...

Reviews: