Hameg HM1008 Manual Download Page 40

40

Subject to change without notice

32.1

 

CH1

If CH1 is selected only CH1 will be turned on, the mode is Yt. Also 
the readout will only display the parameters of CH1. (sensitivity, 
inverted/not inverted, coupling.)
Although CH2 will not appear in the readout it may be used e.g. 
as a trigger input. Its controls are active but are not shown. 

32.2

 

CH2

If CH2 is selected only CH2 will be active, it is Yt mode, and only 
its parameters will be shown in the readout. 
Although CH1 will not appear in the readout it may be used e.g. 
as a trigger input. Its controls are active but are not shown.

32.3.1

 DUAL trace alt./ chop

In dual trace mode both channels are turned on and the parame-
ters of both are shown in the readout. Between the sensitivity indi-
cations there is an indication whether alternate ”alt.“ or chopped 
”chp“ mode is active. Normally, the mode will be automatically set 
by the time base speed selection, but it may be directly set using 
the function pushbutton. For time base speeds of 500 ms/cm to 
500 μs/cm chopped will be used, from 200 μs/cm to 50 ns/cm 
alternate. This refers to unmagnifi ed time bases. 
Alternate is the preferred mode, at any time one channel is dis-
played for a full sweep, after each sweep the other channel has 
its turn. At slow sweep speeds this will cause annoying fl icker, 
at still slower ones the channel switching becomes visible. Here, 
the chopped mode steps in, both channels are switched at some 
high frequency so they are both visible at any sweep speed. 
This is, however, not appropriate for fast sweep speeds as the 
switching may become visible and may interfere with the proper 
signal display. 

32.3.2

 DUAL  (DSO mode).

In DSO mode there is an a/d converter for each channel so both 
are measured simultaneously. Hence no channel switching is 
necessary and no information pertaining to it is shown. 

32.4 ADD

In ADD mode the signals of both channels are algebraically 
added and displayed as one sum signal. The Y position can be 
changed with both position controls. If one channel is inverted 
the difference will be displayed. Only one ”0 V“ symbol will 
be shown in the readout. The ADD mode will be indicated by 
placing a ”+“ symbol between the sensitivity indications of both 
channels. In DSO mode ”1+2“ will be displayed at the end of 
the trace. 
Please note that the results of cursor measurements in this 
mode will only be correct if the sensitivities of both channels are 
identical, otherwise the readout will show ”CH1

<>

CH2“. 

C o n t r o l s  a n d  R e a d o u t          

Automatic voltage measurements can not be performed in ADD 
mode. The readout will show ”n/a“ = not available. 

As the trigger signals are taken off the inputs and not from the 
added signal there is no true reference for the trigger point 
symbol, the symbol will thus be switched off in analog mode. 
However, the LEVEL A/B control 

 is active. 

In DSO mode a trigger time symbol is displayed one line above 
the lowest graticule line to indicate the trigger time position 
along the signal and thus can only be moved horizontally. 

32.5 XY

In this mode CH1 will move the trace in X direction, hence the 
readout will show ”CHX..“, CH2 will move the trace in Y direction, 
hence ”CHY...“ will be shown rather than ”CH2...“.

As the time bases are not involved in XY no time base related 
information will be shown. Also the trigger circuits are disabled 
so no trigger information is shown, either. 

The magnifi er MAG x10 

 is disabled. The ”0-Volt“ symbols 

will be shown as triangles at the right hand graticule and above 
the sensitivities. 

Both the HORIZONTAL 

 or the POSITION 1 

 knobs will 

move the trace horizontally. The Y position is controlled by the 
POSITION 2 knob.

32.5.1

 Analog mode

The CH1 signal can not be inverted, there is hence no menu item 
in the CH1 menu (CH1 pushbutton 

). The TIME/DIV-SCALE-

VAR 

 knob is disabled.

Please note that the bandwidths and phase differences in XY 
analog and DSO modes differ considerably so there may be 
changes in the signal display when switching the mode.

32.5.2

 DSO mode

The readout will indicate the sampling rate with which the a/d 
converters digitise the input signals. The appropriate sampling 
rate must be set depending on the signals and can be selected 
with the TIME/DIV-SCALE-VAR knob, although the time bases 
are disabled. With high sampling rates there may be gaps in 
Lissajous representations. With too low sampling rates the 
display may not allow any more to determine the frequency 
relationship of the signals. 
It is recommended to fi rst look at the signals in DUAL mode 
and to set the sampling rate such that at least one signal period 

VOLTS / DIV

SCALE 

·

 VAR

VOLTS / DIV

SCALE 

·

 VAR

TIME / DIV

SCALE 

·

 VAR

AUTO/

CURSOR

MEASURE

20 V

1 mV

20 V

1 mV

X-POS

LEVEL  A/B

50s

5ns

AUX

HORIZONTAL

DELAY

CH 1/2

CURSOR

MA/REF

ZOOM

MODE

FILTER

SOURCE

TRIG 

d

NORM

HOLD OFF

TRIGGER  

POSITION 1

POSITION 2

CH 1

VERT/XY

CH 2

HOR

MAG

VAR

VAR

VAR

x10

POWER

20

16

19

26

29

18

15

27

23

25

21

28

24

22

30

17

14

13

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Reviews: