Hameg HM1008 Manual Download Page 18

18

Subject to change without notice

internal switching, so the next start is delayed for the so called 
hold-off time, irrespective of the presence of triggers. The 
hold-off time can be extended from its minimum by a factor of 
10:1. Manipulation of the hold-off time and thus of the time for 
a complete sweep period from start to start can be useful e.g. 
when data packets are to be displayed. It may seem that such 
signals can not be triggered. The reason is that the possible 
start of a new sweep does not conincide with the start of a data 
packet, it may start anywhere, even before a data packet. By 
varying the hold-off time a stable display will be achieved by 
setting it just so that the hold-off ends before the start of a data 
packet. This is also handy with burst signals or non-periodic 
pulse trains. 

A signal may be corrupted by noise or hf interference so a double 
display will appear. Sometimes varying the trigger level can not 
prevent the double display but will only affect the apparent time 
relationship between two signals. Here the variable hold-off 
time will help to arrive at a single display. 

Sometimes a double display will appear when a pulse signal 
contains pulses of slightly differing height requiring delicate 
trigger level adjustment. Also here increasing the hold-off 
time will help. 

Whenever the hold-off time was increased it should reset to its 
minimum for other measurements, otherwise the brightness 
will suffer as the sweep rep rate will not be maximum. The 
following pictures demonstrate the function of the hold-off:

Fig. 1: Display with minimum hold-off time (basic setting). 
Double image, no stable display.

Fig. 2: By increasing the hold-off a stable display is achie-
ved.

Time base B (2

nd

 time base). Delaying, Delayed 

Sweep. Analog mode

Consult ”Controls and Readout“ HOR 

 and TIME/DIV. 

 for 

specifi c information.

As was described in ”Triggering and time base“ a trigger will 
start the time base. While waiting for a trigger – after runout 
of the hold-off time – the trace will remain blanked. A trigger 
will cause trace unblanking and the sweep ramp which defl ects 
the trace from left to right with the speed set with TIME/DIV. 
At the end of the sweep the trace will be blanked again and 
retrace to the start position. During a sweep the trace will also 

be defl ected vertically by the input signal. In fact the input signal 
does continuously defl ect the trace vertically, but this will be 
only visible during the unblanking time. This is, by the way, one 
marked difference to DSO operation where the input signal 
is only measured during the acquisition time, for most of the 
time the DSO will not see the signal. Also, in analog mode the 
signal itself will be seen on the screen in real time, whereas 
a DSO can only show a reconstruction of the signal acquired 
some time later. 

In analog mode thus the display will always start on the left. Let 
us assume one period of a signal is displayed at a convenient 
time base setting. Increasing the sweep speed with TIME/DIV. 
will expand the display from the start, so that parts of the signal 
will disappear from the screen. It is thus possible to expand 
the beginning of the signal period and show fi ne detail, but it 
is impossible to show such fi ne detail for ”later“ parts of the 
signal. 

The x10 Magnifi er (MAG x10) may be used to expand the display 
and the horizontal positioning control can shift any part of the 
display into the centre, but the factor of 10 is fi xed.

The solution requires a second time base, called time base B.  

In this mode time base A is called the delaying sweep and 
time base B the delayed sweep. The signal is fi rst displayed 
by TB A alone. Then TB B is also turned on which is the mode
”A intensifi ed by B“. TB B should always be set to a higher sweep 
rate than A, thus its sweep duration will be also shorter than 
that of A. The TB A sweep sawtooth is compared to a voltage 
which can be varied such that TB A functions as a precision 
time delay generator. Depending on the amplitude of the com-
parison voltage a signal is generated anywhere between sweep 
start and end. 

In one of two operating modes this signal will start TB B imme-
diately. The TB A display will be intensifi ed for the duration of 
TB B, so that one sees which portion of the signal is covered by 
TB B, By varying the comparison voltage the start of TB B can 
be moved over the whole signal as it is displayed by TB A. Then 
the mode is switched to TB B. The signal portion thus selected is 
now displayed by TB B. This is called „B delayed by A“. Portions 
of the signal can thus be expanded enormously, however, the 
higher the speed of TB B the darker the display will become as 
the rep rate will remain that of the accepted signal triggers while 
the duration of TB B is reduced with increasing speed. 

In cases where there is jitter the TB B can be switched to wait 
for a trigger rather than starting immediately. When a trigger 
arrives TB B will start on it. The jitter is removed, however, the 
effect is also, that the TB B start now can be only from signal 
period to signal period, no continuous adjustment is possible 
in this mode. 

Alternate sweep

In this mode the signal is displayed twice, with both time bases. 
An artifi cial Y offset can be added in order to separate the two 
displays on the screen. The operation is analogous to Y dual 
trace alternate mode, i.e., the signal is alternately displayed by 
both time bases, not simultaneously which is not possible with 
a single gun crt. What was said above about how TB B can be 
started holds also here. 

T r i g g e r i n g   a n d   t i m e   b a s e          

period

heavy parts are displayed

signal

adjusting

HOLD OFF time

sweep

Fig. 1

Fig. 2

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Reviews: