Hameg HM1008 Manual Download Page 2

2

Subject to change without notice

General information regarding the CE marking

HAMEG instruments fulfi ll the regulations of the EMC directive. The 

conformity test made by HAMEG is based on the actual generic- and 

product standards. In cases where different limit values are applicable, 

HAMEG applies the severer standard. For emission the limits for 

residential, commercial and light industry are applied. Regarding the 

immunity (susceptibility) the limits for industrial environment have 

been used.

The measuring- and data lines of the instrument have much infl uence 

on emmission and immunity and therefore on meeting the acceptance 

limits. For different applications the lines and/or cables used may 

be different. For measurement operation the following hints and 

conditions regarding emission and immunity should be observed:  

1. Data cables

For the connection between instruments resp. their interfaces and 

external devices, (computer, printer etc.) suffi ciently screened cables 

must be used. Without a special instruction in the manual for a reduced 

cable length, the maximum cable length of a dataline must be less than 

3 meters and not be used outside buildings. If an interface has several 

connectors only one connector must have a connection to a cable.

Basically interconnections must have a double screening. For IEEE-bus 

purposes the double screened cables HZ72S and HZ72L from HAMEG 

are suitable.

2. Signal cables

Basically test leads for signal interconnection between test point and 

instrument should be as short as possible. Without instruction in the 

manual for a shorter length, signal lines must be less than 3 meters 

and not be used outside buildings.

Signal lines must screened (coaxial cable - RG58/U). A proper ground 

connection is required. In combination with signal generators double 

screened cables (RG223/U, RG214/U) must be used.

3. Infl uence on measuring instruments

Under the presence of strong high frequency electric or magnetic fi elds, 

even with careful setup of the measuring equipment an infl uence of  

such signals is unavoidable.

Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt
The HAMEG Instruments GmbH herewith declares conformity of the product 
HAMEG Instruments GmbH déclare la conformite du produit

Bezeichnung / Product name / Designation: 
  

 

Oszilloskop

  

 

Oscilloscope

  

 

Oscilloscope

Typ / Type / Type: 

 

HM1008

mit / with / avec:     

Optionen / Options / Options:   

mit den folgenden Bestimmungen / with applicable regulations / avec les 
directives suivantes

 

EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG

 

EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC

 

Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE

Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE

Angewendete harmonisierte Normen / Harmonized standards applied / Normes 
harmonisées utilisées:

Sicherheit / Safety / Sécurité: EN 61010-1:2001 (IEC 61010-1:2001)
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2

Elektromagnetische Verträglichkeit / Electromagnetic compatibility / 
Compatibilité électromagnétique

EN 61326-1/A1 Störaussendung / Radiation / Emission: 
 

Tabelle / table / tableau 4; Klasse / Class / Classe B.

Störfestigkeit / Immunity / Imunitée: Tabelle / table / tableau A1. 

EN 61000-3-2/A14 Oberschwingungsströme / Harmonic current emissions / 
Émissions de courant harmonique: 
 

Klasse / Class / Classe D.

EN 61000-3-3 Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker / 
Fluctuations de tension et du fl icker.

Datum /Date /Date                                          
24. 02. 2005
 

 

 

Unterschrift / Signature / Signatur 

  

 

 

Manuel 

Roth

  

 

 

Manager

 Hersteller 

HAMEG 

Instruments 

GmbH 

KONFORMITÄTSERKLÄRUNG 

 

 

Manufacturer 

Industriestraße 6 

DECLARATION OF CONFORMITY 

 

Fabricant 

D-63533 Mainhausen 

DECLARATION DE CONFORMITE

This will not cause damage or put the instrument out of operation. Small 

deviations of the measuring value (reading) exceeding the instruments 

specifi cations may result from such conditions in individual cases.

4. RF immunity of oscilloscopes.

4.1 Electromagnetic RF fi eld

The infl uence of electric and magnetic RF fi elds may become visible 

(e.g. RF superimposed), if the fi eld intensity is high. In most cases 

the coupling into the oscilloscope takes place via the device under 

test, mains/line supply, test leads, control cables and/or radiation. 

The device under test as well as the oscilloscope may be effected by 

such fi elds.

Although the interior of the oscilloscope is screened by the cabinet, 

direct radiation can occur via the CRT gap. As the bandwidth of 

each amplifi er stage is higher than the total –3dB bandwidth of the 

oscilloscope, the infl uence RF fi elds of even higher frequencies may 

be noticeable.         

4.2 Electrical fast transients / electrostatic discharge

Electrical fast transient signals (burst) may be coupled into the 

oscilloscope directly via the mains/line supply, or indirectly via test 

leads and/or control cables. Due to the high trigger and input sensitivity 

of the oscilloscopes, such normally high signals may effect the trigger 

unit and/or may become visible on the CRT, which is unavoidable. 

These effects can also be caused by direct or indirect electrostatic 

discharge.  

   

HAMEG Instruments GmbH

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...