Hameg HM1008 Manual Download Page 11

11

Subject to change without notice

of the capacitor. It may be assumed that this is negligible for 
frequencies 

>

40 Hz. 

Considering the foregoing you may measure dc signals of up 
to 400 V or pure ac signals of up to 800 V

PP

 with a HZ200 probe. 

Probes with higher attenuation like HZ53 100:1 allow to measure 
dc up to 1200 V and pure ac of up to 2400 V

PP

. (Please note the 

derating for higher frequencies, consult the HZ53 manual). 
Stressing a 10:1 probe beyond its ratings will risk destruction of 
the capacitor bridging the input resistor with possible ensuing 
damage of the scope input! 

In case the residual ripple of a high voltage is to be measured a 
high voltage capacitor may be inserted in front of a 10:1 probe, it 
will take most of the voltage as the value of the probe’s internal 
capacitor is very low, 22 to 68 nF will be suffi cient. 

If the input selector is switched to Ground the reference trace 
on the screen may be positioned at graticule center or else-
where.

DC and ac components of an input signal

The dashed curve shows an ac signal symmetrical to zero. 
If there is a dc component the peak value will be dc + ac peak.

Timing relationships

The repetition frequency of a signal is equal to the number of 
periods per second. Depending on the TIME/DIV setting one or 
more periods or part of a period of the signal may be displayed. 
The time base settings will be indicated on the readout in s/cm, 
ms/cm, μs/cm and ns/cm. Also the cursors may be used to 
measure the frequency or the period.

If portions of the signal are to be measured use delayed sweep 
(analog mode) or zoom (DSO mode) or the magnifi er x 10. Use 
the HORIZONTAL positioning control to shift the portion to be 
zoomed into the screen center.

Pulse signals are characterized by their rise and fall times 
which are measured between the 10 % and 90 % portions. The 
following example uses the internal graticule of the crt, but also 
the cursors may be used for measurement. 

Measurement:

–  Adjust the rising portion of the signal to 5 cm.

–  Position the rising portion symmetrically to the graticule 

centre line, using both Y and X positioning controls.

–  Notice the intersections of the signal with the 10 and 90 % 

lines and project these points to the centre line in order to 
read the time difference. 

In the example it was 1.6 cm at 5 ns/cm equals 8 ns rise time.

When measuring very short rise times coming close to the scope 
rise time it is necessary to subtract the scope’s  (and if used the 
probe’s) rise times geometrically from the rise time as seen on 
the screen. The true signal rise time will become:

t

tot

 is the rise time seen, t

osc

 is the scope’s own rise time 

(3.5 ns with the HM1008), t

t

 is the rise time of the probe, e.g. 

2 ns. If the signal’s rise time is 

>

 34 ns, the rise times of scope 

and probe may be neglected. 

For the measurement of rise times it is not necessary to proceed 
as outlined above. Rise times may be measured anywhere on 
the screen. It is mandatory that the rising portion of the signal 
be measured in full and that the 10 to 90 % are observed. In 
case of signals with over- or undershoot the 0 and 100 % levels 
are those of the horizontal portions of the signal, i.e. the over- 
resp. undershoots must be disregarded for rise and fall time 
measurements. Also, glitches will be disregarded. If signals 
are very distorted, however, rise and fall time measurements 
may be of no value. 

For most amplifi ers, even if their pulse behaviour is far from 
ideal, the following relationship holds:

 350 

350

t

a

  =   ——             B  =   ——  

 B 

t

a

tr/ns  = 350/Bandwidth/MHz  

Connection of signals

In most cases pressing the AUTOSET button will yield a satis-
factory display (see AUTOSET). The following relates to special 
cases where manual settings will be advisable. For a description 
of controls refer to ”Controls and Readout“.

STOP

  Take care when connecting unknown signals to the 

inputs!

It is recommended to use probes whenever possible. Without 
a probe start with the attenuator set to its 20 V/cm position. 
If the trace disappears the signal amplitude may be too large 
overdriving the vertical amplifi er or/and its dc content may be 
too high. Reduce the sensitivity until the trace will reappear 
onscreen. If calibrated measurements are desired it will be 
necessary to use a probe if the signal becomes 

>

160 Vp. Check 

the probe specifi cations in order to avoid overstressing. If the 
time base is set too fast the trace may become invisible, then 
reduce the time base speed. 
If no probe is used at least screened cable should be used, 
such as HZ32 or HZ34. However, this is only advisable for low 

          B a s i c  s i g n a l  m e a s u r e m e n t

t

a

   

8

2

 - 3.5

2

 - 2

2

  =  6.9 ns

t

a

   

t

tot

2

  – t

osc

2

 – t

t

2

voltage

peak

AC

DC

DC

AC

DC + AC

peak

 = 400 V

max

5 cm

t

tot

100%

90%

10%

0%

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Reviews: