Hameg HM1008 Manual Download Page 16

16

Subject to change without notice

without any input signal. Thus there is always a visible trace in 
analog mode, and in DSO mode the trace will also be shown. 
The position of the trace(s) without any signal is then given by 
the settings of the POSITION controls. 

As long as there is a signal scope operation will not need more 
than a correct amplitude and time base setting. With signals 

<

 20 Hz their period is longer than the time the auto trigger 

circuit will wait for a new trigger, consequently the auto trigger 
circuit will start the time base then irrespective of the signal so 
that the display will not be triggered and free run, quite inde-
pendent of the signal’s amplitude which may be much larger 
than the minimum.

Also in auto peak trigger mode the trigger level control is active. 
Its range will be automatically adjusted to coincide with the 
signal’s peak-to-peak amplitude, hence the name. The trigger 
point will thus become almost independent of signal amplitude. 
This means that even if the signal is decreased the trigger will 
follow, the display will not loose trigger. As an example: the 
duty cycle of a square wave may change between 1:1 and 100:1 
without loosing the trigger. 

Depending on the signal the LEVEL A/B control may have to be 
set to one of its extreme positions. 

The simplicity of this mode recommends it for most uncompli-
cated signals. It is also preferable for unknown signals.

This trigger mode is independent of the trigger source and 
usable as well for internal as external triggering. But the signal 
must be 

>

 20 Hz.

Normal trigger mode (See menu MODE)

Consult the chapters: MODE 

 

>

 AUTO, LEVEL A/B 

, FILTER 

 and SOURCE 

 in ”Controls and Readout“. Information 

about how to trigger very diffi cult signals can be found in the 
HOR menu 

 where the functions time base fi ne adjustment 

VAR, HOLD-OFF time setting, and time base B operation are 
explained.

With normal triggering and suitable trigger level setting trigge-
ring may be chosen on any point of the signal slope. Here, the 
range of the trigger level control depends on the trigger signal 
amplitude. With signals 

<

1 cm care is necessary. 

In normal mode triggering there will be no trace visible in the 
absence of a signal or when the signal is below the minimum 
trigger amplitude requirement!

Normal triggering will function even with complicated signals. If 
a mixture of signals is displayed triggering will require repetition 
of amplitudes to which the level can be set. This may require 
special care in adjustment.

Slope selection (Menu FILTER)

After entering FILTER 

 the trigger slope may be selected using 

the function keys. See also ”Controls and Readout“. AUTOSET 
will not change the slope. 

Positive or negative slope may be selected in auto or normal 
trigger modes. Also, a setting ”both“ may be selected which will 
cause a trigger irrespective of the polarity of the next slope. 

Rising slope means that a signal comes from a negative po-
tential and rises towards a positive one. This is independent 
of the vertical position. A positive slope may exist also in the 

negative portion of a signal. This is valid in automatic and 
normal modes.

Trigger coupling (Menu: FILTER)

Consult chapters: MODE 

 

>

 AUTO, LEVEL A/B 

, FILTER 

 

and SOURCE 

 in ”Controls and Readout“. In AUTOSET DC 

coupling will be used unless ac coupling was selected before. 
The frequency responses in the diverse trigger modes may be 
found in the specifi cations.

With internal dc coupling with or without LF fi lter use normal 
triggering and the level control. The trigger coupling selected 
will determine the frequency response of the trigger channel.

AC:

This is the standard mode. Below and above the fall-off of the 
frequency response more trigger signal will be necessary. 

DC:

With direct coupling there is no lower frequency limit, so this 
is used with very slowly varying signals. Use normal triggering 
and the level control. This coupling is also indicated if the signal 
varies in its duty cycle.

HF:

A high pass is inserted in the trigger channel, thus blocking low 
frequency interference like fl icker noise etc. 

Noise Reject:

This trigger coupling mode or fi lter is a low pass suppressing 
high frequencies. This is useful in order to eliminate hf inter-
ference of low frequency signals. This fi lter may be used in 
combination with dc or ac coupling, in the latter case very low 
frequencies will also be attenuated. 

LF:

This is also a low pass fi lter with a still lower cut-off frequency 
than above which also can be combined with dc or ac coupling. 
Selecting this fi lter may be more advantageous than using dc 
coupling in order to suppress noise producing jitter or double 
images. Above the pass band the necessary trigger signal will 
rise. Together with ac coupling there will also result a low 
frequency cut-off. 

Video (tv triggering)

Selecting MODE 

>

 Video will activate the TV sync separator 

built-in. It separates the sync pulses from the picture content 
and enables thus stable triggering independent of the changing 
video content. 

Composite video signals may be positive or negative. The 
sync pulses will only be properly extracted if the polarity 
is right. The definition of polarity is as follows: if the video 
is above the sync it is positive, otherwise it is negative. The 
polarity can be selected after selecting FILTER. If the pola-
rity is wrong the display will be unstable resp. not triggered 
at all as triggering will then initiated by the video content. 
With internal triggering a minimum signal height of 5 mm 
is necessary.

The PAL sync signal consists of line and frame signals which 
differ in duration. Pulse duration is 5 μs in 64 μs intervals. Frame 
sync pulses consist of several pulses each 28 μs repeating each 
half frame in 20 ms intervals.

Both sync pulses differ hence as well in duration as in their 
repetition intervals. Triggering is possible with both.

T r i g g e r i n g   a n d   t i m e   b a s e          

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Reviews: