32
Subject to change without notice
the cursors if the following conditions are met: The cursors
must be activated (Press AUTO/CURSOR-MEASURE pushbutton
>
Cursors
>
Cursors On), press the CH1/2–CURSOR–MA/
REF–ZOOM-pushbutton
select ”Cursors“ or ”Cur. Track“.
The pushbutton will light up blue.
STOP
Note:
The function Cur. Track (cursor tracking) is only
available if 2 cursors are on display. The cursors
will then be moved simultaneously (tracking) wit-
hout changing their respective positions.
CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
This pushbutton calls a menu which allows to select the func-
tions of the controls POSITION 1 and 2. With mathematics
functions this will also affect in some way the SCALE function
which is changed by the VOLTS/DIV controls.
The pushbutton will signal the function activated corresponding
to the front panel labelling:
dark:
Y position CH1 and CH2.
blue:
Y position of cursors.
green:
Y position of:
– Mathematics signal(s)
– Reference signal(s)
– ZOOM or time base B display of signal(s)
VOLTS/DIV–SCALE–VAR knob
This knob is a double function CH1 control.
16.1 Selection of vertical sensitivity
Provided VAR on pushbutton CH1
is not illuminated the sen-
sitivity will be calibrated. Turning the control CCW will decrease,
turning it CW will increase the sensitivity. 1 mV/cm to 20 V/cm
can be selected in a 1-2-5 sequence. The readout will display
the calibrated sensitivity (e.g. ”CH1: 5mV..“). Depending on the
sensitivity selection the signal will be shown with smaller or
greater amplitude.
STOP
Please note:
This sensitivity selection is always active, e.g. also,
if CH2 only was chosen. In that case CH1 may be
used as trigger source.
16.2 Variable control
Select this function with the CH1 pushbutton
>
Variable On.
VAR on the CH1 pushbutton will light up. The readout will add a
C o n t r o l s a n d R e a d o u t
”
>
“ to the sensitivity indication (”CH1
>
5mV..“) in order to show
that the sensitivity is now uncalibrated. The results of cursor
measurements will be fl agged accordingly.
In this mode the sensitivity can be changed with the VOLTS/
DIV–SCALE–VAR control knob from 1 mV/cm to
>
20 V/cm.
16.3
SCALE (DSO mode)
The display height of a mathematic signal resp. a refe-
rence signal can be changed after the function „Math./Ref.“
is selected in the „Pos./Scale“ menu. The latter is called
by pressing the CH1/2–CURSOR–MA/REF–ZOOM-pushbutton
.
VOLTS/DIV–SCALE–VAR knob
This double function control belongs to CH2.
17.1 Selection of vertical sensitivity
The sensitivity will be calibrated provided VAR on the CH2
pushbutton
is not illuminated. Turning the control CCW
will decrease, turning it CW will increase the sensitivity. The
sensitivities can be selected from 1 mV/cm to 20 V/cm in a
1-2-5 sequence. The readout will show the sensitivity (e.g.
”CH2:5mV..“). Depending on the sensitivity the signal will be
displayed with smaller or greater amplitude.
STOP
Please
note:
The sensitivity control is always active, i.e. also if
CH2 is not selected. CH2 may then still be used e.g.
as a trigger source.
17.2 Variable control
The variable control is activated by pressing the CH2 push-
button
>
Variable On. VAR on the pushbutton will light
up. The readout will add the ”
>
“ preceding the sensitivity
(e.g. ”CH2
>
5mV..“) in order to indicate that the sensitivity is
uncalibrated. The results of cursor measurements will be
labelled accordingly.
In this mode the sensitivity may be changed continuously
with the VOLTS/DIV–SCALE–VAR control from 1 mV/cm to
>
20 V/cm.
17.3 SCALE (DSO mode)
The display height of a mathematic signal resp. a refe-
rence signal can be changed after the function „Math./Ref.“
is selected in the „Pos./Scale“ menu. The latter is called
by pressing the CH1/2–CURSOR–MA/REF–ZOOM-push-
button
.
VOLTS / DIV
SCALE
·
VAR
VOLTS / DIV
SCALE
·
VAR
TIME / DIV
SCALE
·
VAR
AUTO/
CURSOR
MEASURE
20 V
1 mV
20 V
1 mV
X-POS
LEVEL A/B
50s
5ns
AUX
HORIZONTAL
DELAY
CH 1/2
CURSOR
MA/REF
ZOOM
MODE
FILTER
SOURCE
TRIG
’
d
NORM
HOLD OFF
TRIGGER
POSITION 1
POSITION 2
CH 1
VERT/XY
CH 2
HOR
MAG
VAR
VAR
VAR
x10
POWER
20
16
19
26
29
18
15
27
23
25
21
28
24
22
30
17
14
13