Hameg HM1008 Manual Download Page 29

29

Subject to change without notice

9.1 Analog and DSO modes

Under ”Save/Recall“ the current instrument settings may be 
saved or settings saved earlier recalled. There are 9 nonvolatile 
memories available.

9.1.1

 Saving the actual settings

Upon pressing the pushbutton the submenu ”Front Panel 
Save“ will be accessed. A memory number will be offered 
(1 to 9) which can be changed by the INTENS knob 

. By pres-

sing ”Save“ all settings will be saved in the memory location 
the number of which was selected.

9.1.2

 Recall the actual settings

In the submenu ”Front Panel Recall“ a memory number will 
be displayed (1 to 9) which can be changed with the INTENS 
knob 

. Pressing the pushbutton ”Recall“ causes loading of 

the settings stored.

9.2

 

DSO mode

The menu options described in 9.1.1 and 9.1.2 are also available 
in DSO mode. Additionally, the menu options ”Reference Save“ 
and ”Reference Display“ will be available. A reference is a signal 
which was resp. can be stored away for later reference to it 9 
nonvolatile memory locations are provided.

9.2.1

 Reference Save

9.2.1.1

 Source x 

In the submenu the source can be selected with the INTENS 
knob 

. Signals from the logic inputs CH3 and CH4 can not be 

stored as references. 

9.2.1.2

 Destination RE x 

There are 9 memory locations available into which reference 
signals from the source selected before can be stored. Use the 
INTENS knob 

 for selection.

9.2.1.3

 Save

Pressing ”Save“ will store the signal from the source selected 
into the memory selected. 

9.2.2

 Reference Display

STOP

 Attention!

 

The display of reference signals automatically swit-
ches previously displayed mathematic signal(s) off 
and vice versa.

9.2.2.1

 RE x, On Off, associated settings

When in this submenu using the INTENS knob  will allow 
selection of 2 reference signals which can then be displayed 
alongside with 2 input signals. 

9.2.2.2

 RE x

After calling this function the memory location can be selected 
with the INTENS knob 

. (RE 1 to 9)

9.2.2.3

 On Off

With the pushbutton on/off control is possible. When the 
contents of the reference memory are displayed, the memory 
number is indicated with RE x (x = 1 to 9) at the right hand 
screen side. Switching to ”on“ will produce another menu item 
(”Assoc.Set“).

STOP

 Please note: 

 

If both reference displays are ”on“ and if both 
memory locations are identical (e.g. RE1, RE1) the 
signal will be displayed twice on the same spot. 

an inadvertent entering of this mode the operation of any control 
will automatically cause this mode to be reset.

As repetitive acquisitions are needed for calculation of an ave-
rage single sweep will not be compatible.

8.4 Roll mode capture/display

Roll mode means that the signal(s) will be continuously acquired 
without the need for a trigger. Hence all controls, displays and 
readouts for the trigger and ZOOM will be disabled. The readout 
will show ”rol“.
The result of the last acquisition will be displayed at the right 
hand edge of the graticule, all formerly acquired signals will 
be shifted one address to the left. The result at the left hand 
screen edge will be dropped. There is no waiting for a trigger 
and thus the hold-off time is minimum. As in any other mode 
the signal acquisition may be stopped and restarted any time 
with the RUN/STOP pushbutton. 
In roll mode the time base available is limited to 50 s/cm to 
50 ms/cm. Faster time bases do not make sense as the signal 
could not be observed any more. 
In case the time base was set outside the limit cited it will be 
automatically changed to the next value within the limits upon 
entering this mode. 

8.5 Peak Detect Auto Off

On Peak Detect Auto mode this mode of acquisition will be au-
tomatically selected if YT and time base speeds of 20 ms/cm to 
2 ms/cm were chosen. This mode is only available with: Refresh, 
Envelope, Average, and Single sweep. The readout will show PD 
preceding the symbol for the operating mode selected. 

At slow sweep speeds the sampling rate will be low, i.e. there 
are relatively large gaps between samples. Glitches occurring 
during those gaps will go by unnoticed. With Peak Detect, how-
ever, the highest sampling rate will be used such that glitches 
will be caught. The samples will be looked at and the ones with 
the highest values stored and displayed.

8.6 Random Auto Off

Provided single sweep was not selected Random Sampling will 
be automatically selected beginning at a certain sweep speed. 
The time base setting will be indicated in the readout, e.g. ”RS: 
10 GSa“ (= Random sampling with 10 GS/s effective sampling 
rate), the real time time base speed will be 5 ns/cm. Without 
Random Sampling ”RS“ Real Time Sampling will be used with 
a maximum sampling rate of 1 GS/s (one channel only) or 
500 kSa/s (two channel mode).

Random Sampling requires repetitive signals, each signal peri-
od will contribute one sample. At an effective sampling rate of 
10 GS/s the time difference from sample to sample along the 
signal period will be 0.1 ns. However, note that with random 
sampling the samples are not taken in sequence along the sig-
nal period but randomly  with respect to it. Random Sampling 
allows to generate the 200 points per cm in X direction at the 
fastest time base of 5 ns/cm. 

Remark:

 5 ns/cm are also available in other modes. In Real 

Time Sampling mode and 1 GS/s on one channel each 1 ns a 
sample is taken, hence at 5 ns/cm there are 5 points per cm. 
The ”missing“ 195 points are generated by interpolation using 
sin x/x. 

  SAVE/RECALL

This pushbutton will open up a menu. The number of choices 
in this menu is dependent upon whether there is analog or 
DSO mode. 

          C o n t r o l s  a n d  R e a d o u t

Summary of Contents for HM1008

Page 1: ...100 MHz CombiScope HM1008 Manual English...

Page 2: ...MC Directive 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Dire...

Page 3: ...Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode See menu MODE...

Page 4: ...me Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect RS 232 Interface optional USB RS 232 IEEE 488 Ethernet USB Signal display Yt and XY Interpolation Sinx...

Page 5: ...e 50 ppm Display 1 MEMORY ZOOM max 50 000 1 Bandwidth X Amplifier 0 100 MHz 3 dB X Y phase shift 3 100 MHz Digital Storage Sampling rate real time 2x 500 MSa s 1 GSa s interleaved Sampling rate random...

Page 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Page 7: ...xplosion The operating position may be any however suf cient ventila tion must be ensured convection cooling Prolonged operation requires the horizontal or inclined position STOP Do not obstruct the v...

Page 8: ...tes the current function of POSITION 1 and 2 controls VOLTS DIV SCALE VAR knob 32 Channel 1 Y de ection coef cient Y variabel and Y scaling setting VOLTS DIV SCALE VAR knob 32 Channel 2 Y de ection co...

Page 9: ...Component Tester Left socket is galvanically connected with protective earth MEMORY oom COMPONENT TESTER PROBE ADJ C O M B I S C O P E POWER 38 39 40 EXIT MENU VOLTS DIV SCALE VAR VOLTS DIV SCALE VAR...

Page 10: ...ich of course must have a suf cient dc rating Care must be taken however when charging and discharging a large capacitor Dc coupling is preferable with all signals of varying duty cyc le otherwise the...

Page 11: ...ead the time difference In the example it was 1 6 cm at 5 ns cm equals 8 ns rise time When measuring very short rise times coming close to the scope rise time it is necessary to subtract the scope s a...

Page 12: ...specially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops may exist Also most instruments wi...

Page 13: ...herefore please observe the following rule Always look at the two signals in the one channel only or the dual modes and make sure that they are within the permissible input signal range this is the ca...

Page 14: ...ator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the calculation to angles 90 degrees This is wh...

Page 15: ...e time base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the...

Page 16: ...ve of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vertical position A positive slope ma...

Page 17: ...ed with SOURCE Alt 1 2 The read out will display Tr alt but no more the trigger point symbol indicating level and time position Instead an arrow pointing upwards will indicate the trigger time positio...

Page 18: ...time whereas a DSO can only show a reconstruction of the signal acquired some time later In analog mode thus the display will always start on the left Let us assume one period of a signal is displayed...

Page 19: ...n this mode the Y ampli ers and the time base are turned off Only individual components may be tested i e they must not be part of a circuit if voltages are to be applied to the BNC connectors If the...

Page 20: ...conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a plain short although the part is good With enhancement type MOSFETs...

Page 21: ...t the signal is a sine wave It is easily under stood that in order to depict an unknown signal shape one needs at least 1 or 2 points per centimeter in other words the useful signal frequency is only...

Page 22: ...al capture is triggered in SINGLE REFRESH ENVELOPE and AVERAGE modes and untriggered in ROLL and XY modes The normal Refresh mode is similar to the operating mode of an analog scope Triggering will ca...

Page 23: ...lained the maximum sampling rate must be reduced for slow time base settings This may cause aliases If e g a sine wave is sampled only with one sample per period and if it should be synchronous with t...

Page 24: ...eters are N 8 2 no parity 8 bits data 2 stop bits RTS CTS hardware protocol Selection of Baud rate Baud rate setting is automatic Range 110 to 115200 no parity 8 bits data 2 stop bits The rst SPACE CR...

Page 25: ...ess and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in the readout may disappear this will reappear immediately upon leaving the menu Each...

Page 26: ...displayed by ZOOM digital mode only RO Int Readout intensity Focus Focus for signal and readout Readout On Off Turning the readout off will eliminate interference of the readout with the signal s The...

Page 27: ...s blue The operating mode Yt or XY will not be affected If component test mode was selected possible only in analog mode and the scope is switched to digital the operating mode last used when in DSO m...

Page 28: ...ntre With the 2nd acquisition the display will start at the screen left In most cases this is meaningless but the scope may seem not to react at slow time base settings combined with long Posttrigger...

Page 29: ...without the need for a trigger Hence all controls displays and readouts for the trigger and ZOOM will be disabled The readout will show rol The result of the last acquisition will be displayed at the...

Page 30: ...1 MByte samples may be acquired but only up to 2 KBytes per channel can be displayed hence possibly min or max values may not be shown Otherwise the display is as described above AUTOSET Choosing AUT...

Page 31: ...Mathematics signal position DSO mode only The POSITION 1 control will assume the function of position con trol for mathematics signals after the following procedure Press the MATH pushbutton Display u...

Page 32: ...alibrated The results of cursor measurements will be agged accordingly In this mode the sensitivity can be changed with the VOLTS DIV SCALE VAR control knob from 1 mV cm to 20 V cm 16 3 SCALE DSO mode...

Page 33: ...AUTO CURSOR MEASURE pushbutton Pressing this pushbutton will open the menu Measurement which offers the submenus Cursors and Auto If the submenu Cursors was selected and a measuring mode Cursors On m...

Page 34: ...n mode the level can only be selected between the signal s peak values The movement of the trigger symbol is vertical only The range of this symbol is limited in order to prevent that this symbol will...

Page 35: ...ode of the scope C o n t r o l s a n d R e a d o u t any level can be set If the trigger level is set such that no trig gers are generated the automatic triggering will nevertheless start the time bas...

Page 36: ...nly This display will light up if the hold off time was set to 0 in order to indicate that the longer than minimum hold off time may cause a lower rep rate of the time base and thus a darker display S...

Page 37: ...the readout Z and is calibrated Turning the knob CCW will decrease turning it CW will increase the time base speed This can be selected from 20 ms cm to 5 ns cm in a 1 2 5 sequence The maximum expansi...

Page 38: ...ded to the time base B display Press the CH1 2 CURSOR MA REF ZOOM pushbutton which calls the Pos Scale menu Press the function push button TB B this will cause the POSITION 1 knob to act as the positi...

Page 39: ...between 1 mV cm to 20 V cm VERT XY pushbutton This pushbutton switches the vertical menu on off This menu allows to select the operating modes of the vertical ampli ers C o n t r o l s a n d R e a d o...

Page 40: ...correct if the sensitivities of both channels are identical otherwise the readout will show CH1 CH2 C o n t r o l s a n d R e a d o u t Automatic voltage measurements can not be performed in ADD mode...

Page 41: ...n XY DSO mode both channels may be inverted 32 6 Bandwidth Full 20 MHz This pushbutton will select full or 20 MHz bandwidth Full Full bandwidth will be the one given in the speci cations 20 MHz Provid...

Page 42: ...l s a n d R e a d o u t PROBE ADJ connector A square wave signal of 0 2 Vpp is available for the adjustment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and callin...

Page 43: ...43 Subject to change without notice C o n t r o l s a n d R e a d o u t...

Page 44: ...ruments GmbH Industriestra e 6 A Rohde Schwarz company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales...

Reviews: