Temperature
Compensation
Automatic from –10 to 105 °C
(14.0 to 221 °F) with selection
for NTC 300 ohm thermistor, Pt
1000 ohm RTD, or Pt 100 ohm
RTD temperature element, or
manually fixed at a user-entered
temperature; additional
selectable temperature
correction factors (ammonia,
morpholine, or user-defined
pH/°C linear slope) available for
pure water automatic
compensation from 0.0 to 50 °C
(32 to 122 °F)
Automatic from –10 to 105 °C
(14.0 to 221 °F) with selection
for NTC 300 ohm thermistor, Pt
1000 ohm RTD, or Pt 100 ohm
RTD temperature element, or
manually fixed at a user-entered
temperature; additional
selectable temperature
correction factors (ammonia,
morpholine, or user-defined
pH/°C linear slope) available for
pure water automatic
compensation from 0.0 to 50 °C
(32 to 122 °F)
N/A
Measurement
Accuracy
±0.02 pH
±0.02 pH
±5 mV
Temperature Accuracy
±0.5 °C (0.9 °F)
±0.5 °C (0.9 °F)
±0.5 °C (0.9 °F)
Repeatability
±0.05 pH
±0.05 pH
±2mV
Sensitivity
±0.01 pH
±0.01 pH
±0.5 mV
Calibration Methods
Two point automatic, one point
automatic, two point manual,
one point manual.
Two point automatic, one point
automatic, two point manual,
one point manual.
one point manual
Maximum Probe
Immersion Depth/
Pressure
Submersible to 107 m (350
ft)/1050 kPa (150 psi)
Immersion only
Submersible to 107 m (350
ft)/1050 kPa (150 psi)
Sensor Interface
Modbus
Modbus
Modbus
Probe Cable Length
6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension for
analog sensor with digital
gateway
10 m (31 ft) for sensor with
integral digital electronics
6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension for
analog sensor with digital
gateway
10 m (31 ft) for sensor with
integral digital electronics
6 m (20 ft) + 7.7 m (25 ft)
interconnect cable extension
for analog sensor with digital
gateway
10 m (31 ft) for sensor with
integral digital electronics
Probe Weight
316 g (11 oz)
870 g (31 oz)
316 g (11 oz)
Probe Dimensions
See
Figure 2 on page 9
through
Figure 3 on page 9
.
See
Figure 4 on page 9
.
See
Figure 2 on page 9
through
Figure 3 on page 9
.
1
Most pH applications are in the 2.5 to 12.5 pH range. The pHD™ Differential pH sensor with the wide-range glass process electrode
performs exceptionally well in this range. Some industrial applications require accurate measurement and control below 2 or above 12 pH.
In these special cases, please contact the manufacturer for further details.
2
For best ORP measuring results in solutions containing zinc, cyanide, cadmium or nickel, the manufacturer recommends using the pHD™
ORP sensor equipped with a gold electrode.
3
PEEK
®
is a registered trademark of ICI Americas, Inc.
4
Ryton
®
is a registered trademark of Phillips 66 Co.
5
Kynar
®
is a registered trademark of Pennwalt Corp.
6
Viton
®
is a registered trademark of E.I. DuPont de N Co.
Table 2 Digital Gateway Specifications
Weight
145 g (5 oz)
Dimensions
17.5 x 3.4 cm (7 x 1
3
/
8
in.)
Operating Temperature
–20 to 60 °C (–4 to 140°F)
Table 1 Differential pH and ORP Sensor Specifications (continued)
Specification Category
pH Sensors
1
Stainless Steel pH Sensor
ORP Sensors
2
Summary of Contents for sc200
Page 5: ...DOC023 53 80040 sc200 Controller 01 2014 Edition 7 User Manual ...
Page 6: ......
Page 10: ...Table of Contents 4 ...
Page 20: ...14 nglisK ...
Page 23: ... nglisK 17 ...
Page 53: ......
Page 56: ...Visit us at www hach com ...
Page 74: ...20 Installation ...
Page 90: ...Visit us at www hach com Maintenance ...
Page 96: ...Visit us at www hach com ...
Page 100: ...Visit us at www hach com ...
Page 106: ...Visit us at www hach com ...
Page 108: ...Visit us at www hach com ...
Page 110: ...Visit us at www hach com ...
Page 115: ......
Page 148: ...Figure 4 5 Transistor Outputs 4 8 ...
Page 176: ...Figure 8 1 Dimensions 8 2 ...
Page 195: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 265: ...Figure 1 1 Typical ModBUS Network Topography 1 3 ...
Page 267: ...Figure 1 2 Typical Communications Wiring Scheme 1 5 ...
Page 268: ...Figure 1 3 Signal Boost Sensor Diagram 1 6 ...
Page 270: ...Figure 2 1 Typical Communication Cable Wiring for Signal Boost Option 2 2 ...
Page 271: ...Figure 2 2 Sensor Display Screens 2 3 ...
Page 290: ......
Page 292: ......
Page 348: ...PAGE 54 APPENDIX D MOUNTING 15 APPENDIX D MOUNTING Figure 9 Mounting Detail ...
Page 351: ......
Page 352: ...Powerware Series Eaton 9130 UPS 700 3000 VA User s Guide ...
Page 472: ... 0 5 3 5 6 A B 3 4 ñ 4 9 0 ñ ñ ñ 2 ñ 5 7 8 1 5 CB N4 C N4 9 2 2 6 9 E 3 7 2 5 0 0 2 3 2 ...
Page 473: ... 0 5 3 5 6 A B 3 4 ñ 4 9 0 ñ ñ ñ 2 ñ 5 5 3 0 44 7 3 7 4 2 K 3 50 0 ...
Page 476: ......
Page 477: ......
Page 479: ......
Page 555: ......
Page 556: ......
Page 557: ......
Page 558: ......
Page 559: ......
Page 560: ......
Page 561: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 562: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 563: ...UNIT HEATERS FOR STEAM OR HOT WATER Rosemex Products HEATING 01 2003 ...
Page 571: ......
Page 579: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 584: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 Ƭ ...
Page 594: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ȋ Ȍ ...
Page 624: ......
Page 625: ......
Page 626: ......
Page 627: ......
Page 628: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 636: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 637: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 638: ......
Page 639: ......
Page 640: ......
Page 641: ......
Page 642: ......
Page 643: ......
Page 644: ......
Page 645: ......
Page 646: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ǡ ǡ ...
Page 650: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 4 M5 9 1 2 3 4 ...
Page 651: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 5 ...
Page 653: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 7 3 3 7 7 3 3 ...
Page 654: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 8 1 2 4X 8X 1X ...
Page 672: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 805: ...Notes ENGLISH ...
Page 806: ...Notes ENGLISH ...
Page 807: ...Notes ENGLISH ...
Page 809: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 810: ......
Page 811: ......
Page 812: ......
Page 861: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 862: ......
Page 863: ......
Page 864: ......
Page 865: ......
Page 866: ......
Page 867: ......
Page 868: ......
Page 869: ......
Page 870: ......
Page 871: ......
Page 872: ......
Page 873: ......
Page 874: ......
Page 875: ......