![Hach sc200 User Manual Download Page 21](http://html.mh-extra.com/html/hach/sc200/sc200_user-manual_576786021.webp)
Table 1 AC power wiring information (AC powered models only)
Terminal Description
Color
²
North
America
Color
²
EU
1 Hot
(L1)
Black
Brown
2 Neutral
(N)
White
Blue
—
Protective Earth (PE)
Ground lug
Green
Green with yellow
stripe
Table 2 DC power wiring information (DC powered models only)
Terminal Description
Color
²
North
America
Color
²
EU
1 +24
VDC
Red
Red
2
24 VDC return
Black
Black
—
Protective Earth (PE)
Ground lug
Green
Green with yellow
stripe
Alarms and relays
The controller is equipped with four unpowered, single pole relays rated
100-250 VAC, 50/60 Hz, 5 amp resistive maximum. Contacts are rated
250 VAC, 5 amp resistive maximum for the AC powered controller and
24 VDC, 5A resistive maximum for the DC powered controller. The
relays are not rated for inductive loads.
Wiring relays
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the
instrument when making electrical connections.
W A R N I N G
Potential fire hazard. The relay contacts are rated 5A and are not
fused. External loads connected to the relays must have current
limiting devices provided to limit current to < 5 A.
W A R N I N G
Potential fire hazard. Do not daisy-chain the common relay
connections or jumper wire from the mains power connection inside
the instrument.
W A R N I N G
Potential electrocution hazard. In order to maintain the NEMA/IP
environmental ratings of the enclosure, use only conduit fittings and
cable glands rated for at least NEMA 4X/IP66 to route cables in to the
instrument.
AC line (100
²
250 V) powered controllers
The wiring compartment is not designed for voltage connections in
excess of 250 VAC.
24 VDC powered controllers
W A R N I N G
Potential electrocution hazard. AC mains powered controllers
(115 V–230 V) are designed for relay connections to AC mains circuits
(i.e., voltages greater than 16 V-RMS, 22.6 V-PEAK or 35 VDC).
W A R N I N G
Potential electrocution hazard. 24 V powered controllers are designed
for relay connections to low voltage circuits (i.e., voltages less than
16 V-RMS, 22.6 V-PEAK or 35 VDC).
(
nglis
K
15
Summary of Contents for sc200
Page 5: ...DOC023 53 80040 sc200 Controller 01 2014 Edition 7 User Manual ...
Page 6: ......
Page 10: ...Table of Contents 4 ...
Page 20: ...14 nglisK ...
Page 23: ... nglisK 17 ...
Page 53: ......
Page 56: ...Visit us at www hach com ...
Page 74: ...20 Installation ...
Page 90: ...Visit us at www hach com Maintenance ...
Page 96: ...Visit us at www hach com ...
Page 100: ...Visit us at www hach com ...
Page 106: ...Visit us at www hach com ...
Page 108: ...Visit us at www hach com ...
Page 110: ...Visit us at www hach com ...
Page 115: ......
Page 148: ...Figure 4 5 Transistor Outputs 4 8 ...
Page 176: ...Figure 8 1 Dimensions 8 2 ...
Page 195: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 265: ...Figure 1 1 Typical ModBUS Network Topography 1 3 ...
Page 267: ...Figure 1 2 Typical Communications Wiring Scheme 1 5 ...
Page 268: ...Figure 1 3 Signal Boost Sensor Diagram 1 6 ...
Page 270: ...Figure 2 1 Typical Communication Cable Wiring for Signal Boost Option 2 2 ...
Page 271: ...Figure 2 2 Sensor Display Screens 2 3 ...
Page 290: ......
Page 292: ......
Page 348: ...PAGE 54 APPENDIX D MOUNTING 15 APPENDIX D MOUNTING Figure 9 Mounting Detail ...
Page 351: ......
Page 352: ...Powerware Series Eaton 9130 UPS 700 3000 VA User s Guide ...
Page 472: ... 0 5 3 5 6 A B 3 4 ñ 4 9 0 ñ ñ ñ 2 ñ 5 7 8 1 5 CB N4 C N4 9 2 2 6 9 E 3 7 2 5 0 0 2 3 2 ...
Page 473: ... 0 5 3 5 6 A B 3 4 ñ 4 9 0 ñ ñ ñ 2 ñ 5 5 3 0 44 7 3 7 4 2 K 3 50 0 ...
Page 476: ......
Page 477: ......
Page 479: ......
Page 555: ......
Page 556: ......
Page 557: ......
Page 558: ......
Page 559: ......
Page 560: ......
Page 561: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 562: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 563: ...UNIT HEATERS FOR STEAM OR HOT WATER Rosemex Products HEATING 01 2003 ...
Page 571: ......
Page 579: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 584: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 Ƭ ...
Page 594: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ȋ Ȍ ...
Page 624: ......
Page 625: ......
Page 626: ......
Page 627: ......
Page 628: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 636: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 637: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 638: ......
Page 639: ......
Page 640: ......
Page 641: ......
Page 642: ......
Page 643: ......
Page 644: ......
Page 645: ......
Page 646: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ǡ ǡ ...
Page 650: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 4 M5 9 1 2 3 4 ...
Page 651: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 5 ...
Page 653: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 7 3 3 7 7 3 3 ...
Page 654: ... 2014 Pentair Equipment Protection PH 763 422 2211 hoffmanonline com 87563688 8 1 2 4X 8X 1X ...
Page 672: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 805: ...Notes ENGLISH ...
Page 806: ...Notes ENGLISH ...
Page 807: ...Notes ENGLISH ...
Page 809: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 810: ......
Page 811: ......
Page 812: ......
Page 861: ... ǁǁǁ ǁŝůĚƐƚŽŶĞ ĐŽŵ 2 0DLQ 6WUHHW 3K 3HQWLFWRQ D DQDGD 9 ...
Page 862: ......
Page 863: ......
Page 864: ......
Page 865: ......
Page 866: ......
Page 867: ......
Page 868: ......
Page 869: ......
Page 870: ......
Page 871: ......
Page 872: ......
Page 873: ......
Page 874: ......
Page 875: ......