3.6.2 Analog output signal
The analog output signal, representing the measured gas concentration, is available from the
instrument in the ranges from 0 to 20 mA, or in the range from 4 to 20 mA. In addition, the model
3662EX has a second analog output of 0 to 2 Volts. Depending on the instrument model, separate
output connections are available for the "safe" area and for the "hazardous" area.
Scale these outputs between two custom concentration limits. If the lower of these limits is
represented by AL, and the upper one by AH, the relation between the output current I (or voltage V)
and the gas concentration C is as follows:
0 – 20 mA
C = AL + (AH - AL) * I/20
4 – 20 mA
C = AL + (AH - AL) * (I - 4)/16
0 – 2 Volts
C = AL + (AH - AL) * V/2
The analog output saturates at 20 mA (or 2 Volts), or is blocked at 0/4 mA (or 0 Volts) under various
conditions, as summarized below:
Analog output
Explanation
0/4 mA (or 0 Volts)
Thermal cutoff or below analog output low limit (AL)
0/4 mA (or 0 Volts) < signal < 20 mA
Normal operating condition
20 mA (or 2 Volts)
Over range, above analog output high limit (AH), or sensor disconnected
3.6.3 Alarm relay output (3660EX only)
There are three separate alarm relay connections on the instrument's output card (NO and NC), for
each alarm limit relay (High and Low), and system (SYST) alarm. The following table lists the
possible alarm conditions.
Conditions
Low alarm
NC to Com*
High alarm
NC to Com*
Low alarm
NO to Com*
High alarm
NO to Com*
Alarm syst
NC to Com*
Alarm syst
NO to Com*
Instrument power
switch off
closed
closed
open
open
closed
open
No alarm:
Low Limit < measured
gas < High Limit
open
open
closed
closed
open
closed
Low alarm:
measured gas < Low
Limit
closed
open
open
closed
open
closed
High alarm:
measured gas > High
Limit
open
closed
closed
open
open
closed
System test
closed
open
Sensor not connected
closed
open
open
closed
closed
open
Thermal cut off
closed
open
* Low NC, High NC, Low NO, High NO, and Com identify the alarm relay connections on the instrument's output
card.
Note: The system alarm is turned on when the following situations are met:
Instrument off - Alarm high - Alarm low - Sensor out - Thermal cut off
English
9
Summary of Contents for ORBISPHERE 366 Series
Page 268: ...3 3 3 4 Orbisphere 3660EX 3662EX 3660EX 3662EX EN 61326 1 3662EX EN61000 4 5 32956 C 24 48 268...
Page 272: ...3 6 3 3660EX SYST 4 EN 61010 1 3660EX 272...
Page 276: ...200 150 276...
Page 279: ...8 3662EX 1 6 2 7 PG 3 8 4 O2 O3 H2 9 J11 5 Lemo 10 10 PE 279...
Page 295: ...Range 0 1 5 10 4 0 100 8 8 Measurements View 8 9 Sensor Current View 295...
Page 327: ......