Figure 7 3660EX instrument with external pressure sensor (Example is wired RS 232)
1
Connection to "hazardous area": Inside the
instrument, cables to "hazardous area" must never
overlap the "safe area".
8
Current output to hazardous area
2
Connections to "safe area": "Hazardous area"
cables must be shielded up to the connector, then
the shield must be connected to ground.
9
External pressure sensor
3
Color code for sensor cable item no. 32514E
10
Current output to safe area and alarm outputs
4
Shield ground
11
RS-232 or RS-485
5
Lemo 10
12
115 or 230 VAC (L= Line, N=Neutral)
6
PG gland
13
PE-Earth
7
O2, O3 or H2 sensor (hazardous area)
14