![Hach ORBISPHERE 366 Series Basic User Manual Download Page 11](http://html1.mh-extra.com/html/hach/orbisphere-366-series/orbisphere-366-series_basic-user-manual_3152369011.webp)
Figure 4 3662EX installation
1
Safe area
7
Optional external pressure sensor (32 570E.wxyz)
2
Hazardous area
8
Sensor cable (32 514E.mm) 4+shield
3
Safety barrier
1
: 32956.A 110–230 VAC; 32956.C
24–48 VDC
9
0/4–20 mA shielded cable, Rload 100 Ohm max.,
compliance 2 V & 0–2V
4
Power supply shielded cable
10
Depending of barrier and cables characteristics
5
Sensor cable (i32 515E.mm) 10+shield
11
Repeater 4/20 mA (32954)
6
Sensor head (31xxxE or GA2800 ATEX)
Note: The 3662EX may be installed with a different configuration and/or accessories.
4.2 Waterproof cable gland fitting
The nickel-plated cable glands are EMC-type, and are designed to ground the cable shield to the
instrument metal case.
1
All the devices below are accessory equipments
English
11
Summary of Contents for ORBISPHERE 366 Series
Page 268: ...3 3 3 4 Orbisphere 3660EX 3662EX 3660EX 3662EX EN 61326 1 3662EX EN61000 4 5 32956 C 24 48 268...
Page 272: ...3 6 3 3660EX SYST 4 EN 61010 1 3660EX 272...
Page 276: ...200 150 276...
Page 279: ...8 3662EX 1 6 2 7 PG 3 8 4 O2 O3 H2 9 J11 5 Lemo 10 10 PE 279...
Page 295: ...Range 0 1 5 10 4 0 100 8 8 Measurements View 8 9 Sensor Current View 295...
Page 327: ......