49
Section 7
Faults, causes, rectification
Note: FILTRAX has three different types of error messages
7.1 Messages
Only text is displayed referring to the need for the
replacement of consumables - the green signal lamp
remains active.
7.2 Warnings
Text is displayed, both signal lamps are active and the
warning relay contact is set. If possible rectify without delay!
FAULT
CAUSE
RECTIFICATION
AIRFILTER COMP.
The operating time counter for the compressor air
filter [AIR F.C.] has expired
Replace compressor air filter and reset operating
time counter
AIRFILTER CASE
The operating time counter for the control unit air
filter [AIR F.H.] has expired
Replace control unit air filter and reset operating time
counter
PUMPWHEELS
The operating time counter for the pump rollers
[P.ROLL] has expired
Replace pump rollers and reset operating time
counter
PUMPCASSETTE
The operating time counter for the pump
cartridges [P.CARTR.] has expired
Replace pump cartridges and reset operating time
counter
COMPRESSOR
The operating time counter for the compressor
[COMP.] has expired
Replace compressor and reset operating time
counter
FAULT
CAUSE
RECTIFICATION
PUMPTUBES
The operating time counter for the pump tubes
[P.TUBES] has expired
Replace pump tubes and reset operating time
counter
LESS SAMPLE
The amount of sample from a module is below the
threshold set under [WARNING < ]
Check all tubes, if necessary clean filter modules
TEMPERATURE
The temperature in the control unit is too high, the
filter module venting is shut down automatically
Check fans in the control unit, if necessary replace
filter mats, reduce ambient temperature
TEST SETTINGS
On the [+DEVICEDATA] menu some settings may
have been lost
Check all instrument settings on the [+DEVICEDATA]
menu and [+SERVICE] menu
Summary of Contents for Filtrax
Page 2: ......
Page 4: ...2 Contents Section 8 Spare parts 51 Section 9 Warranty and liability 53 Section 10 Contact 55 ...
Page 6: ...4 Technical data ...
Page 12: ...10 General information Fig 2 6 Filter module ...
Page 16: ...14 Installation Fig 3 9 Module carrier and filter module ...
Page 31: ...Installation 29 ...
Page 33: ...Commissioning 31 Fig 4 23 Inserting filter modules ...
Page 35: ...Commissioning 33 ...
Page 43: ...Operation 41 ...
Page 55: ...52 ...
Page 57: ...54 Warranty and liability ...