11
Section 3
Installation
D A N G E R
Only qualified experts should conduct the tasks described in this section.
D A N G E R
Select an appropriate installation location for the instrument.
Plan out the mechanical mount before positioning poles or drilling holes. Make sure the
mount has a sufficient bearing capacity. The dowels must be selected and authorized
according to the condition of the wall.
The manufacturer shall accept no liability if the instrument is installed incorrectly.
Plan how to lay cables and tubes and their path in advance. Lay the tubes, data cables
and power cables without any bends and so they do not pose a tripping risk.
Do not connect the electrical supply to the mains until the instrument is completely wired
and protected against short circuits.
Sufficiently protect the electrical power supply against short circuits.
For the external power supply, always connect a residual-current circuit breaker (trip
current max.: 30 mA) between the mains and the system.
If the instrument is to be installed outdoors, connect a surge arrester between the mains
and the system.
Products intended by the manufacturer for outdoor use offer a higher level of protection
against the penetration of liquids and dust. If these products are connected to a mains
outlet with a cable and plug rather than a permanently connected cable, the plug and
outlet are much more susceptible to liquid and dust penetration. The operator must
sufficiently protect the plug and outlet against liquid and dust penetration in accordance
with local safety regulations. If the instrument is to be used outdoors, it must be
connected to a suitable outlet with a protection type of at least IP44 (splash protection).
WA R N I N G
Electrical dangers and fire hazard. Use only the supplied power cable.
Only qualified experts may perform the tasks described in this section of the manual, in
compliance with all locally applicable safety regulations.
N O T I C E
Protect the device against extreme temperatures from heaters, direct sunlight and other
heat sources.
C A U T I O N
Note the weight (control unit approx. 22 kg, module carrier with 5 m suction tube approx.
9 kg) of the instrument. Do not try to carry the instrument without help. Use only suitable
lifting devices for the transport.
Summary of Contents for Filtrax
Page 2: ......
Page 4: ...2 Contents Section 8 Spare parts 51 Section 9 Warranty and liability 53 Section 10 Contact 55 ...
Page 6: ...4 Technical data ...
Page 12: ...10 General information Fig 2 6 Filter module ...
Page 16: ...14 Installation Fig 3 9 Module carrier and filter module ...
Page 31: ...Installation 29 ...
Page 33: ...Commissioning 31 Fig 4 23 Inserting filter modules ...
Page 35: ...Commissioning 33 ...
Page 43: ...Operation 41 ...
Page 55: ...52 ...
Page 57: ...54 Warranty and liability ...