26
Installation
Fig. 3-20
Terminal assignments
Table 3-1 Terminal assignments
1.
Mains cable (power supply)
3.
Heater connecting cable, sample tube
2.
Earth connection
4.
Heater connecting cable, suction tube
Sicherungen/Fuses
250 V
F1:
F2:
F3:
F4:
T3,15A L
T3,15A L
T6, 3A L
T6, 3A L
T4A E
T4A E
T7A E
T7A E
IEC
UL
N O T I C E
Route the signal lines (21, 22, 23, 24) separated from the power cable.
Terminal no.
Connection
1
N (230 V AC / 50-60 Hz, option: 115 V AC)
2
L (230 V AC / 50-60 Hz, option: 115 V AC)
7
N (sample tube heater)
8
N (suction tube heater)
9
L (sample tube heater)
10
L (suction tube heater)
18
Earth for screening signal wires
19
Earth for suction tube heater
20
Earth for sample tube heater
21/22
Floating fault signalling contacts (normally open)
23/24
Floating warning contacts (normally open)