Operation
5.5 [+OP.COUNTERS] menu
The [+OP.COUNTERS] menu provides information on the duration of the use of
the various instrument components and consumables. It is not needed for
operation and the instrument settings.
Note:
While the instrument is in this menu, normal operation continues. If a key is not
pressed for longer than ten minutes, the display automatically returns to the display of the
data on operation.
*If replacement date passed.
MENU
Command
Description
+OP.COUNTERS
CLEANED
Month when cleaning was last performed, is automatically updated during
[+FILTERCLEANING]
MODUL A
Operating time for filter module A in days
MODUL B
Operating time for filter module B in days
P.TUBES IN
Remaining operating time for both pump tubes in days (91 - 0 - negative number
*
)
P.CARTR.IN
Remaining operating time for both pump cartridges in days (365 - 0 - negative number
*
)
P.ROLL.IN
Remaining operating time for pump rollers in days (365 - 0 - negative number
*
)
AIR F.C.IN
Remaining operating time for the compressor air filter in days (365 - 0 - negative number
*
)
AIR F.H.IN
Remaining operating time for the two air filters in the control unit in days (91 - 0 - negative number
*
) - it is not
necessarily imperative to change the air filter on the air outlet side so frequently
COMP. IN
Remaining operating time for the compressor in days (730 - 0 - negative number
*
)
Summary of Contents for Filtrax
Page 2: ......
Page 4: ...2 Contents Section 8 Spare parts 51 Section 9 Warranty and liability 53 Section 10 Contact 55 ...
Page 6: ...4 Technical data ...
Page 12: ...10 General information Fig 2 6 Filter module ...
Page 16: ...14 Installation Fig 3 9 Module carrier and filter module ...
Page 31: ...Installation 29 ...
Page 33: ...Commissioning 31 Fig 4 23 Inserting filter modules ...
Page 35: ...Commissioning 33 ...
Page 43: ...Operation 41 ...
Page 55: ...52 ...
Page 57: ...54 Warranty and liability ...