operation
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 34 of 216
5.4.4
Switch off procedure model saturn 450/300
no.
Illustration
Description
[01]
Press the “power off” push button.
[02]
Switch off the running machine by using the main
switch at the right side of the external UserTerminal
(control unit). The main switch has to be turned 90°
counterclockwise as shown in the picture.
The main switch can be secured against
unauthorized access with a padlock.
N
The ON-/OFF intervals must not be shorter than 1-2 minutes (for models with 3-phase
connection: 2-3 minutes). This may otherwise lead to interference and/or failure of the backup
and/or data loss and/or data corruption.
N
Models for medical application with the isolation transformer have an inrush current limiter
(surge guard). Too short ON/OFF intervals will lead to a deactivation of the inrush current limiter
and result in an overload of the circuit fuse.
N
If the device is used daily, we recommend switching it on in the morning and leave it in stand-by
mode during the day.
N
Press the emergency stop only in case of danger.
N
The emergency stop is not to be used as a normal stop key.
5.5
Emergency Stop, stop buttons, stop functions and safety features
In case of emergency, e.g. danger of stumbling and/or falling, etc., grab the front crossbar and/or both side handrails
and jump onto the foot rails (step platforms) with both feet.
The emergency stop button is to be pressed immediately!
h/p/cosmos is in compliance with the norm EN 957-6 and offers more safety features than required, as shown in the
following tables. In all below mentioned cases, all displayed values will be reset to “0”.
N
Use the emergency stop only if in danger.
N
The emergency stop is not to be used as a normal stop key.