optional settings: administrator options
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 153 of 216
OP48
Maximum speed during maximum
activation of the frequency inverter
(applicable for digital RS485 and/or
analogue control 0…10 volt of the
frequency inverter)
The maximum frequency inverter speed is achieved if the MCU
(within UserTerminal) sets the control voltage (analogue voltage for
frequency inverter) at 10 V. This means maximum frequency of the
inverter.
The maximum speed during maximum frequency of the inverter is set
with the input of the correct device type at option number 03 (OP03)
of the administrator options. Option 48 is necessary for manual speed
calibration or special speed transformation or drive motors (special
speeds) only.
When changing OP 48, always pay attention to the correct setting of
the OP 05 (maximum speed), OP 11 (speed measurement; has to be
switched on) and OP 34 (speed calibration value).
Examples for the different special speeds and the appropriate
maximum inverter frequencies are to be found in the list in chapter
10.1 entitled “List of running machines types”
N
Make sure that the correct device type is set at administrator OP03
N
Go to administrator OP 05 and set maximum possible speed.
N
Set the device in the manual mode to the maximum possible speed
by using key "+".
N
Check the analogue voltage at the MCU-FU control with a volt
meter. It should be 10 V. Make a note of the voltage in the service
report (not possible with digital control via RS485)!
N
Read the speed and make a note of the value in the h/p/cosmos
service report: OP 48: _____ km/h
N
Stop the running belt with the STOP key.
N
Go to administrator OP 48 and insert the previously measured
value in km/h (maximum possible speed at full FU-activation).
N
Go again to administrator OP 05 and indicate the desired /
appropriate maximum speed.
Info 1: If no speed scan (speed sensor) is set under OP11, the speed
calibration value will be calculated automatically with OP48.
Info 2: Even if no speed scan is set, the speed calibration value is of
internal significance. In that case the scanning impulses (actual
impulse) are deduced or simulated directly from the analogue voltage
and not measured via the speed sensor input.
Range of adjustment: 1.00 to 99.99 km/h
Fixed values selectable with up- /down-key:
20; 22; 22.5; 26; 30; 30.3; 40; 40.9; 41.2; 45.3; 46; 50; 60; 80
OP49
Type of frequency inverter (FI)
control
Setting of FI control. Two types are available for selection:
§ $ %
Analogue 0 to 10 V
AnA
LOG
Digital via RS-485
rS-
485